[ViZion DR+ 1417V2 Wireless Panel Calibration Manual]
[Ultra 4.2.9.0]
12
Customer Support
1.800.366.5343
–
The Calibration wizard will open into a new
window, presenting the
Welcome
screen.
3. Set X-ray environment:
•
Tube SID
=
50” or higher (at least
greater than 47”).
•
No Collimation
= open collimator
blades at least 1” past the edges of
the panel on all sides.
•
No Grid
= remove any grid or
pressure grid cover.
•
No Objects
= remove any objects
from the surface of the panel.
4. Click the arrow in the top right corner
labeled
Start the journey of
Calibration
.
5. Click
Skip
, in the top right corner, at the
Offset Calibration
screen.
6. Click
Skip
, in the top right corner, at the
Generator Parameter Learning
screen
2.2.3 Panel Gain Calibration
Note
: the Gain Calibration screen
requires 5 exposures. The left-hand
side of the screen tells you the
suggested kV and mAs to use for
the current exposure. The right-
hand side of the screen is divided
into 5 blank gain exposure frame
icons. Each exposure frame is
numbered sequentially, with the kV
to use noted next to each one. A
green box surrounds the current
exposure frame.
1. Set your techniques to
70kV / 4.7 mAs
and take your
first
exposure.