[ViZion DR+ 1417V2 Wireless Panel Calibration Manual]
[Ultra 4.2.9.0]
13
Customer Support
1.800.366.5343
–
Note
: after taking an exposure, the
current gain exposure frame will still
be surrounded in green. It will either
have a green checkmark, if it was
successful, or a yellow exclamation
point, if not within 100 +/- of the
expected center average of 7000.
2. If the exposure was NOT acceptable
(yellow status), adjust mAs up or down,
click
Prep
, and take another exposure.
3. If the exposure was acceptable (green
status), click
Next
to move to the next
gain frame.
4. After taking all 5 gain exposures, click
Back
if you want to retake any previous
exposures, then click
Prep
and retake
the exposure. Otherwise, if you are
satisfied with all exposures, click
Next
in
the top right corner.
5. Wait for the
Gain Map Generation
to
complete and then click
Next
, in the top
right corner.
Note
: Click
Next
in the top right corner,
and proceed to
Section 2.2.4
for
Panel Defect
Calibration
,
ONLY if
needed
. A defect calibration is only
needed if you are seeing dead
pixels or lines in your images.
Otherwise proceed to Section
2.2.5
to download the new Gain
calibration file to the panel.