Regulatory Information
KONICA MINOLTA,
INC..……………..…………………………………………………………………………………………………
……………….………..83
Electrical
fast
transients/burst
IEC 61000-4-4
±2 kV
100kHz repetition frequency
Conducted
disturbances
induced
by RF fields
IEC 61000-4-6
3V, 0.15MHz-80MHz
6V in ISM bands between 0.15MHz and
80MHz
80%AM at 1kHz
Signal input/output parts Port
Phenomenon
Basic EMC
standard
Immunity test levels
Professional healthcare facility
environment
Electrostatic
Discharge
IEC 61000-4-2
±8 kV contact
±2kV, ±4kV, ±8kV, ±15kV air
Electrical
fast
transients/burst
IEC 61000-4-4
±1 kV
100kHz repetition frequency
Conducted
disturbances
induced
by RF fields
IEC 61000-4-6
3V, 0.15MHz-80MHz
6V in ISM bands between 0.15MHz and
80MHz
80%AM at 1kHz
The following shows information on reference cables provided against EMC
Cable
Recommended
cable length
Shielded
or
Unshielded
Number
Cable classification
AC Power Cable
3m
Unshielded
1 pcs
AC Power
DC Power Cable
3.5m
Unshielded
1 pcs
DC Power
LAN Cable
(configuration mode)
3m
Shielded
1 pcs
Signal
Important information regarding Electromagnetic Compatibility (EMC)
SKR 4000 requires special precautions regarding EMC and needs to be installed only by KONICA or
authorized personnel and put into service according to EMC information provided in the user manual.
SKR 4000 in use may be susceptible to electromagnetic interference from portable and mobile RF
communications such as mobile (cellular) telephones. Electromagnetic interference may result in
incorrect operation of the system and create a potentially unsafe situation.
SKR 4000 conforms to this EN60601-1-2:2015 standard for both immunity and emissions.
Nevertheless, special precautions need to be observed: