23
WARNING
Do not disassemble or modify the CM-3700A-U or AC adapter.
Doing so may cause a fire or electric shock.
The CM-3700A-U should not be operated if it is damaged, or if smoke or strange odors occur.
Doing so may result in a fire. In such situations, turn the power OFF immediately, disconnect the
AC adapter from the AC outlet, and contact the nearest Konica Minolta-authorized service facility.
CAUTION
When cleaning, disconnect the power plug. Connecting may cause electric shock.
Be careful around openings in the CM-3700A-U. Failure to do so
may result in fingers being trapped causing injury.
Zero Calibration Box and White Calibration Plate
Wipe gently with a dry soft cloth. If dirt is difficult to remove, dampen a cloth with commercially available lens
cleaning liquid and wipe. Then wipe off the liquid with a cloth dampened with water, and leave it to dry.
{
When cleaning, take care not to scratch the zero calibration box or white calibration plate.
Target Mask
Use a blower to remove dirt and dust from the target masks.
{
Do not touch the black-coated surface of the target masks with fingers or wipe it with a cloth. If the target masks
get so dirty that dirt cannot be removed using a blower, contact the nearest Konica Minolta-authorized service
facility.
Inside Integrating Sphere
1.
Set the specular component to SCI using the
software.
2.
Open the Measurement port baseplate and
remove dust and dirt using a blower.
{
Do not touch the white-coated inner surface of the
integrating sphere, wipe it with a cloth or place any
object against it. If the white-coated surface gets
so dirty that dirt cannot be removed using a blower,
contact the nearest Konica Minolta-authorized
service facility.
Cleaning the CM-3700A-U and Accessories