3.2 Connecting System
3-4
3.2.2 Setting Up Network
Connect the REGIUS console to the unit one to one and configure settings for network connection.
*: Connection using cross cables (without using an Ethernet hub) is possible.
Requirements
The procedures described in this manual require the following tools:
Important
When Connecting to an Existing REGIUS Console
For establishing temporary connection with the unit, the settings for the REGIUS console must
be reset to the factory settings (1:1 connection and post-registration).
Back up the setup information (system and JM information) of the REGIUS console, then
change the settings as described below.
•
Network address setting screen
IP address: 192.168.20.90; Subnet mask: 255.255.255.0
•
JOBM INFO screen
Select Built-in Job Manager and Post-registration.
•
CCU INFO screen
Host name: R170-0001; Host name (on the Job Manager): r170-00001;
IP address: 192.168.20.170
•
PostgreSQL access tool
c_status screen: Set to CS#-0001 only;
r_status screen: Set to r170-0001 only;
relations screen: Set to Only one relationship between CS#-0001 and r170-0001;
sys_config screen: Set reg_type to1
(CS#: CS1 for CS-3, CS2 for CS-2)
For information about the procedure for changing the REGIUS console settings, refer to the
REGIUS console Installation/Service Manual.
No.
Tools
Remarks
1
Ethernet cable (2)
Used to connect the REGIUS console to the unit. A cross
cable can be used instead.
Use a category 5e or higher cable.
2
Ethernet hub
3
Maintenance PC
Used as an alternative to the REGIUS console. The
REGIUS console application must be installed.
Ethernet hub
REGIUS console
*