The maintenance procedures that must be performed by the user are:
a) cleaning: wash after each use with lukewarm drinking water (max 40 ° C), possibly with
the addition of a neutral detergent (e.g. Marseille soap). Rinse and leave to dry in the
shade, away from direct heat,
b) disinfection, when deemed necessary or after prolonged inactivity (more than 21 days),
preceded by cleaning: wipe with a clean cloth , soaked in a water solution of sodium
hypochlorite (bleach) witha 0,1% concentration (1000 ppm). When blood or other
bodily fluids are present, the recommended concentration of sodium hypochlorite is
0,5%(5000ppm).
Repairs may only be carried out by the manufacturer.
The user is permitted to replace with new, original parts only those parts
mentioned in paragraph 3.4.2.
The “
SHELL
” stretcher is made from materials with high resistance to wear and to external
agents. Despite this, practical conditions of use will require some maintenance to be carried
out and, in certain special cases, even some repairs.
MAINTENANCE
AND REPAIR
5.1
GENERAL
5.2
MAINTENANCE
5.3
REPAIR
CHAPTER
5
Maintenance and repairs must be recorded in an appropriate log, an example
of which is given in chapter 10.