MJ400252C1
- 13 -
Description of mode No. and set values
Mode No.
Damping constant setting
P
-
01
A
B
C
D
1.
0
Set range: 0.0
-
25.0 s
Mode No.
Unit of flow rate setting and full scale flow rate setting
P
-
02
Unit of flow rate setting
A
B
C
D
3
1
・・・
mL/s
2
・・・
mL/min
3
・・・
L/min
4
・・・
L/h
Full scale flow rate setting
A
B
C
D
0 5.
0
0
Input range: mL/s
・・・
3.14
-
99.99
mL/min
・・・
188
-
9999
L/min
・・・
0.18
-
11.30
L/h
・・・
11
-
678
※
After a full scale flow rate has been set, when
YES is selected with the up key
[
∧
], the value will be registered. If
NO is selected, the value will not be
changed and the indicator will return to mode No. indication.
Mode No.
K factor value setting
P
-
03
A
B
C
D
1. 0 0 0
Set range:
:
0.500
-
5.000
Mode No.
Low flow rate cut setting
P
-
04
A
B
C
D
0 0
Set range: 0
-
20%