K71000 Automatic PMCC Flash Point Analyzer
Operation and Instruction Manual
K71000-Manual
-
35-
12. Factory Settings -
The Factory Settings Menu can only be accessed when logged into the Analyzer
under the
Factory
Login. The Factory settings menu is shown in
Figure 25
below. In the Factory
Settings Menu, the user or administrator is capable of adjusting parameters such as the Dip
Mechanism, Lift Mechanism and Electrical Igniter Intensity. Please note that these parameters have
been preset at the Koehler Factory for Optimal performance of the Analyzer.
a. Adjust Dip Mechanism:
Pressing the
Up
and
Down
Arrows will adjust the level at which the
igniter will dip into the shutter assembly during testing. The
Find Dip Home
button will
automatically bring the Dip Mechanism to the Home Position (Highest Position), acting as a reset
for making future adjustments. The
Check Dip
Button will activate the Dip Mechanism for a Test
Dip. The
Dip
On/Off Switch will activate or deactivate the dipping mechanism.
b. Adjust Lift Mechanism:
Pressing the
Up
and
Down
Arrows will adjust the level at which the Lift
Mechanism will be in relation to the analyzer base. The
Find Lift Home
button will bring the lift to
the Home (Lowest Position). The
Test Position
Button will bring the lift to the Test Position (Low
Position on Test Cup). The
Change Position
Button will bring the lift to the change position
(Middle). The
Clean Position
button will bring the lift to the Clean Position (Upper). The
Stop Lift
button will automatically stop lift from moving.
c. Set Electrical Igniter Intensity:
The Intensity of the Electrical Ignitor can be adjusted as a
percentage both for when the Igniter is intensifying for a
Test Dip
as well as for when the Igniter
is being used to
Re-light
the gas flame in the event the flame goes out while using gas ignition.
Enter the desired intensity from (0 - 100) % and click the
Save
button. The Electrical and Gas
Igniter can also be turned Off or On by pressing the toggle switch shown.
Figure 25: Factory Settings Screen