K71000 Automatic PMCC Flash Point Analyzer
Operation and Instruction Manual
K71000-Manual
-
17-
4.3
Cover Assembly
Figure 3: Instrument Descriptions_Cover Assembly
23. Stirring Propeller:
To maintain sample homogeneity throughout the duration of the test.
24. Sample Temperature Probe:
Measures sample temperature during testing.
25. Ionization Detector:
For automatic detection of the sample's flash point.
26. Flame Detection Thermocouple:
For continuous monitoring of pilot flame during testing. In the
event the test flame goes out, the Flame detection thermocouple will automatically detect this and
initiate re-lighting of the flame.
27. Pilot Flame:
The pilot flame ensures the test flame is lit prior to each test dip.
28. Electric Ignitor:
Alternative means for applying heat source to initiate a flash point. The electric
ignitor is also used to light the pilot flame.
29. Gas Ignitor:
Referee means for applying heat source to initiate a flash point.
30. Application Mechanism:
Houses gas / electric ignitor assembly. Automatically lowers ignitor into
test cup at intervals in accordance to the prescribed test method.
31. Electric Ignitor Cable:
Reinforced cable for connecting to Ignitor Connection Port
(Figure 1, Item 8)
.
32. Gas Tubing:
Provides means for gas source to reach gas ignitor.