Koehler K44100 Operation And Instruction Manual Download Page 4

- 3 - 

1 Introduction 

 

The Automatic Non-Woods Metal Noack 

Evaporative Apparatus tests for the evaporation loss 
tendencies of lubricating oils at temperatures of up to 
275°C.  The newly designed electrically heated 
aluminum block allows for testing without the use of 
hazardous Woods Metal. The Noack tester is 
equipped with an Electronic regulator allowing for 
automatic control of temperature and differential 
pressure. The system is managed by an integrated 
6.5” Touch Screen Panel PC by means of the Noack 
Evaluation Software run by a Windows® based 
operating System. The Evaluation Software is 
capable of recording all analytical parameters, 
allowing for user customizable parameters, methods 
and result reports as well as printing graphs and test 
results. 

 

• Conforms to ASTM D5800, Procedure B 
• 6.5” Integrated Touch Screen Panel PC 
• Integrated Vacuum Pump with automatic electronic 

control system 

• Direct sample temperature measurement via 

PT100 probe 

• Equipped with high resistant Kalrez valve, inlet filter 

to remove product residuals 

• USB port for connection to an external printer 

and/or external PC 

• Storage capacity for more then 60,000 analysis 
• CE Marked 
 

This manual provides important information 

regarding safety, technical references, installation 
requirements, operating condition specifications, 
user facility resource requirements, and operating 
instructions for the Automatic Non-Woods Metal 
Noack Evaporative Apparatus. This manual should 
also be used in conjunction with applicable published 
laboratory procedures. Information on these 
procedures is given in section 1.2. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

1.1  Koehler’s Commitment to Our 

Customers 

 

Providing quality testing instrumentation and 

technical support services for research and testing 
laboratories has been our specialty for more than 
50 years. At Koehler, the primary focus of our 
business is providing you with the full support of 
your laboratory testing needs. Our products are 
backed by our staff of technically knowledgeable, 
trained specialists who are experienced in both 
petroleum products testing and instrument service 
to better understand your requirements and 
provide you with the best solutions. You can 
depend on Koehler for a full range of accurate and 
reliable instrumentation as well as support for your 
laboratory testing programs. Please do not 
hesitate to contact us at any time with your 
inquiries about equipment, tests, or technical 
support. 
 

Toll Free: 1-800-878-9070 (US only) 

Tel: +1 631 589 3800 • Fax: +1 631 589 3815 

Email: [email protected] 

http://www.koehlerinstrument.com 

 

1.2  Recommended Resources and 

Publications

 

 

1.  American Society for Testing and Materials 

(ASTM) 
100 Barr Harbor Drive 
West Conshohocken, Pennsylvania 19428-
2959, USA 
Tel: +1 610 832 9500 • Fax: +1 610 832 9555 
http://www.astm.org • email: [email protected] 

 

ASTM Publication: 

 

  ASTM D5800: Standard Test Method 

Evaporation Loss of Lubricating Oils by 
the Noack Method

 

 

 
2.  Energy Institute (IP) 

61 New Cavendish Street 
London, WIM 8AR, United Kingdom 
Tel: 44 (0)20 7467 7100  
Fax: 44 (0)20 7255 1472 
http://www.energyinstpubs.org.uk/ 

 

IP Publication: 

 IP 

421 

 
 
 
 
 
 

Summary of Contents for K44100

Page 1: ...mia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Analysis Instrumen...

Page 2: ...1...

Page 3: ...Packing List 5 3 2 Unpacking 5 3 3 Installation 5 3 4 Additional Accessories Required for Running Tests 5 3 5 Optional Accessories Available 5 4 Descriptions 6 4 1 Piping Diagram 6 4 2 Instrument Cont...

Page 4: ...s manual should also be used in conjunction with applicable published laboratory procedures Information on these procedures is given in section 1 2 1 1 Koehler s Commitment to Our Customers Providing...

Page 5: ...to access the analyzer STAY ALERT Do not operate this unit while under the influence of medication alcohol or drugs Equipment Modifications and Replacement Parts Any modification or alteration of thi...

Page 6: ...shipping container and equipment Do not return goods to Koehler without written authorization 3 3 Installation Equipment Placement Place the instrument on a firm surface in a secure area 1 Connect th...

Page 7: ...6 4 Descriptions 4 1 Piping Diagram Figure 1 Piping Diagram...

Page 8: ...K441XX Automatic Non Woods Metal Noack Evaporative Apparatus Operation and Instruction Manual K441XX Manual 7 4 2 Instrument Controls and Connections Figure 2 Top Front Left Side View...

Page 9: ...K441XX Automatic Non Woods Metal Noack Evaporative Apparatus Operation and Instruction Manual K441XX Manual 8 Figure 3 Top Left Back Side View Figure 4 Analyzer Complete with Assembled Glassware Set...

Page 10: ...bration date Figure 7 Calibration Screen 4 On the left side of the screen a graph shows the calibration curve along with the selected probe s value in both C and volts 5 On the right side of the scree...

Page 11: ...emperature Control via PT100 probe in the Heating Block Bath Method B Temperature Control via PT100 probe in the Crucible 4 Select the set point temperature for the test based on the selected method S...

Page 12: ...egulate the aspiration vacuum to 2 mbar by slowly closing the bleed valve halfway and regulating the vacuum regulator on the analyzer s side in order to reach 20 0 2mm on the inclined manometer 2mbar...

Page 13: ...for 30 minutes NOTE WHILE REMOVING THE CRUCIBLE AND PLACING IT IN THE WATER BATH MAKE SURE THE SAMPLE DOES NOT SPLASH THE CRUCIBLE S COVER 7 After 30 minutes carefully dry the outside of the crucible...

Page 14: ...e operator to select the sample o Press the Load Analysis Button in between the two arrows to view the analysis Figure 13 Results Browser Main Screen If the final weight has not been introduced a ques...

Page 15: ...hler Representative as each printer has different characteristics and specific drivers Touching the Show Details button causes the analysis to reload and the data to be displayed second by second on t...

Page 16: ...test is run clean the Woullf bottles and pipes Each time a test is run clean the heating jacket Periodically clean the vacuum filter 7 2 Replacement Parts Some instrument parts may need to be replaced...

Page 17: ...rize the return of the product to the factory The sole responsibility of Koehler Instrument Company and the purchaser s exclusive remedy for any claim arising out of the purchase of any product is the...

Page 18: ...K441XX Automatic Non Woods Metal Noack Evaporative Apparatus Operation and Instruction Manual K441XX Manual 17 Notes...

Page 19: ...K441XX Automatic Non Woods Metal Noack Evaporative Apparatus Operation and Instruction Manual K441XX Manual 18 Notes...

Page 20: ...K441XX Automatic Non Woods Metal Noack Evaporative Apparatus Operation and Instruction Manual K441XX Manual 19 Notes...

Reviews: