K430X2 Dual Twin Foaming Characteristics Test Bath 

Operation and Instruction Manual 

 

K430X2-Manual

    

 

-

10-

 

 

  Temperature Range: 5 to 40°C 

 

  Elevation to 2000 meters 

 

  Relative Humidity: < 80% 

 
Ventilation.

  A  fume  hood  or  exhaust  system  is 

required  when  operating  the  unit.  Flammable 
vapors  and/or  steam  are  generated  during 
operation  and  must  not  be  permitted  to 
accumulate.  A  canopy-style  hood  may  be  used  if 
the height from the top of the unit to the canopy is 
5  feet  or  less.  The  exhaust  blower  should  have  a 
rating of 1000 C.F.M. or greater. 
 

Power. 

Connect  the  line  cords  to  properly  fused 

and grounded receptacles with the correct voltage 
as  indicated  in  section  1.3  or  on  the  back  of  the 
unit.  
 

WARNING:

 

For  safety,  disconnect  the 

power  when  performing  any  maintenance  and/or 
cleaning.

 

Do 

NOT

  turn  the  power  on  unless  the 

bath  is  filled  with  the  proper  medium;  otherwise, 
damage  may  occur  to  the  unit  and  the  warranty 
will be void. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

4  Descriptions

 

 

4.1  Main Instrument Controls 

 

 

 

Figure 1: Main Instrument Descriptions 

 

1.  Power Switch.

 This switch controls the power 

to motor and heater / temperature controller of 
its corresponding side. When the power switch 
is  in  the 

ON

  position,  the  digital  temperature 

controller and the stirrer are powered on. 

 

2.  Temperature  Controller.

  The  temperature 

controller  regulates  the  bath  temperature  its 
corresponding  bath.  Refer  to  Section  4.3  for 
full operational details. 

 

3.  Flowmeters. 

To  regulate  the  flow  of  air  into 

each air delivery tube assembly.  

 

4.  Stirrer Motor.

 The stirrer constantly circulates 

bath medium to prevent temperature gradients 
and  ensures  temperature  stability.  When 
cleaning  and/or  servicing,  please  be  sure  to 
disconnect unit power to avoid possible injury. 

 

5.  Heat  Shield.

 

An 18” safety  shield  is provided 

to  protect  the  user  from  accidentally  touching 
the “Hot” Bath while operating the instrument.  

 

6.  Temperature  Probe.

  Detects  the  bath 

temperature  and  communicates  with  the 
temperature  controller  to  maintain  desired 
temperature for each bath.  

 

7.  Air Delivery Tube.

 Connect to air conditioning 

coil and flowmeter for passing air through gas 

Summary of Contents for K430X2

Page 1: ...ohemia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Analysis Instrumentation Custom Design Manufacturing K430X2 Dual Twin Foaming Characteristics Test Apparatus ...

Page 2: ......

Page 3: ...with the applicable standards set forth in this certification Specifications ASTM D892 IP 146 DIN 51566 FTM 791 3211 FTM 791 3213 NF T 60 129 This unit is tested before it leaves the factory to ensure total functionality and compliance to the above specifications and ASTM standards Test and inspection records are on file for verification Jesse Kelly Application Engineer Koehler Instrument Company ...

Page 4: ......

Page 5: ...K430X2 Dual Twin Foaming Characteristics Test Bath Operation and Instruction Manual K430X2 Manual 1 ...

Page 6: ...K430X2 Dual Twin Foaming Characteristics Test Bath Operation and Instruction Manual K430X2 Manual 2 ...

Page 7: ...asic information on subsequent processing or recycling prior to disposal of the equipment http www dpa system dk en WEEE Products Private or professional use Responsibility for Registration and Annual Reporting Koehler will not sell directly to end users in the EU and so has no responsibility to register within each EU state and to make annual reports Koehler declares that this responsibility is b...

Page 8: ...eparately collected WEEE Mercury containing components such as switches or backlighting lamps compact fluorescent lamps CFL Batteries Printed circuit boards if the surface of the printed circuit board is greater than 10 square centimeters about 4 sq in Toner cartridges liquid and pasty as well as color toner Chlorofluorocarbons CFC hydrochlorofluorocarbons HCFC or hydrofluorocarbons HFC hydrocarbo...

Page 9: ... Flowmeter Setup and Operation 11 4 3 Flowmeter Calibration Data 11 4 4 Temperature Controller Operation 11 5 Operation 13 5 1 Bath 13 5 2 Power 13 5 3 Preparation of Glassware and Tubing 13 6 Safety Features 14 6 1 Over Temperature Protection 14 6 2 Over Power Protection 14 7 Maintenance 14 7 1 Routine Maintenance 14 7 2 Bath Cleaning 14 7 3 Replacement Parts 15 8 Wiring Diagrams 16 8 1 Unit Wiri...

Page 10: ...K430X2 Dual Twin Foaming Characteristics Test Bath Operation and Instruction Manual K430X2 Manual 6 12 Warranty 17 13 Returned Goods Policy 18 Notes 19 ...

Page 11: ...specialists who are experienced in both petroleum products testing and instrument service to better understand your requirements and provide you with the best solutions You can depend on Koehler for a full range of accurate and reliable instrumentation as well as support for your laboratory testing programs Please do not hesitate to contact us at any time with your inquiries about equipment tests ...

Page 12: ... and Replacement Parts Any modification or alteration of this equipment from that of factory specifications is not recommended voids the manufacturer warranty product safety performance specifications and or certifications whether specified or implied and may result in personal injury and or property loss Replacement parts must be O E M exact replacement equipment Over Temperature Protection This ...

Page 13: ...rom box to cart and from cart to bench Individuals must lift in accordance to proper technique See Figure below 7 Lift instrument from cart and place on bench 8 Ensure that all parts listed on the packing list are present Inspect the unit and all accessories for damage If any damage is found keep all packing materials and immediately report the damage to the carrier We will assist you with your cl...

Page 14: ...1 Main Instrument Controls Figure 1 Main Instrument Descriptions 1 Power Switch This switch controls the power to motor and heater temperature controller of its corresponding side When the power switch is in the ON position the digital temperature controller and the stirrer are powered on 2 Temperature Controller The temperature controller regulates the bath temperature its corresponding bath Refe...

Page 15: ...the duration specified in the test procedure and disconnect the tygon tubing from the flowmeter outlet when the aeration period as elapsed 4 3 Flowmeter Calibration Data K43002 0 12 Air Flow Rate 94 mL min Float Material Sapphire Float Density 3 98 g ml Metered Fluid Air Minimum Flow 16 4 ml min Maximum Flow 159 0 ml min Tolerance 2 F S Scale Reading mm Flowrate ml min 65 159 0 60 139 0 55 121 2 5...

Page 16: ...is a value observed in the upper red LED display add it to the calculated difference obtained in the previous step This is the offset value c Press the Program Key The lower green display will flash Use the up or down keys to adjust to the new calibration offset value on the upper red display calculated in the previous step When the value has been entered the controller will automatically store th...

Page 17: ...s that will decompose or volatilize producing flammable vapors below the maximum instrument operating temperature 25 C NOTE Coloration of the Mineral or Silicon Bath Oil is an indication that the Medium must be changed See Draining instructions in section 7 3 of this document When a decision is made to discard this material dispose of this material and its container to a hazardous of special waste...

Page 18: ...e servicing to avoid exposure to high voltages and or temperatures which may result in personal injury or death If you have any questions about maintaining your equipment then please do not hesitate to contact the Koehler technical service department 7 1 Routine Maintenance The K430X2 Dual Twin Foaming Test Bath requires little routine maintenance to provide many years of continuous service Howeve...

Page 19: ...0Hz 7A 278 001 002 Fuse 1A Slo Blo 5 X 20 mm 332 001 003 12 x 18 Borosilicate Bath Jar K23425 03000 18 Safety Shield K23700 03014A Motor 230V 50 60HZ 1 15HP K430 0 13 Air Outlet Elbow Tube K430 0 8 Rubber Stopper K430 3A Air Inlet Delivery Tube without diffuser K43002 0 12 Low Flowmeter K43002 0 13 Test Tube Rack K43002 0 15 Locking Post Assembly K43012 1000ml Cylinder w Ring K43002 0 9 Hot Bath O...

Page 20: ...K430X2 Dual Twin Foaming Characteristics Test Bath Operation and Instruction Manual K430X2 Manual 16 8 Wiring Diagrams 8 1 Unit Wiring ...

Page 21: ... Dual Twin Foaming Characteristics Test Bath should not require service Any service problem can be quickly resolved by contacting Koehler s technical service department either by letter phone fax or email In order to assure the fastest possible service please provide us with the following information Model Number Serial Number Date of Shipment 11 Storage This laboratory test instrument is equipped...

Page 22: ...er the item s to be returned and the reason for the return You will be issued a Returned Authorization RA number which must be prominently displayed on the shipping container when you return the material to our plant Shipping containers without an RA number prominently displayed with will be returned to the sender Goods must be returned freight prepaid Returns will be subject to a restocking charg...

Page 23: ...K430X2 Dual Twin Foaming Characteristics Test Bath Operation and Instruction Manual K430X2 Manual 19 Notes ...

Page 24: ...K430X2 Dual Twin Foaming Characteristics Test Bath Operation and Instruction Manual K430X2 Manual 20 Notes ...

Reviews: