![Koehler K41100 Operation And Instruction Manual Download Page 10](http://html1.mh-extra.com/html/koehler/k41100/k41100_operation-and-instruction-manual_1984936010.webp)
K41100
2 - TECHNICAL ASPECT
2.1 - Scope of application
This appliance is intended for the treatment of samples in compliance with the
following standards:
ASTM D 4530
2.2 - Technical characteristics
Ambient temperature for use
15 to 30° C
Ambient relative humidity for use
Max 85% non-condensing
Oven temperature range
Ambient T° to 550° C
Oven temperature probe
Thermocouple J
Resolution of temperature measurement 1° C
Type of oven
Conform to the standard ASTM D 4530
Cooling of oven
Accelerated by injection of compressed
air for forty minutes
Gas connections at the rear of the
Instant connections dia. 4 mm for
appliance
nitrogen and dia. 8 mm for compressed
air
Maximum upstream gas pressure
2 bar
Type of gas authorised
Air for cooling the oven - nitrogen for the
oven
Maximum sample load
12 vials each of 2 ml
Maximum heating kinetic
15° C/min
Use voltage
230 V 50/60 Hz
Power
1500 W
Temperature safety device
By separate J thermocouple set at
550° C
Electrical safety of electronic unit
By fuse (10 A)
Supply voltage
230 V - 50 Hz
Dimensions (L x W x H)
250 x 300 x 600 mm
Weight
19 kg