Koehler K39400 Operation And Instruction Manual Download Page 16

K394XX Water Separability Tester 

Operation and Instruction Manual 

 
 

K394XX-Manual

    

 

-

12-

 

 

 

 

Figure 11 . Stirrer Not Engaged 

 
 

 

 

Figure 12. Stirrer Engaged 

 

 

 

Figure 13: Run Screen 

 
5.  After pressing the 

START TEST

 button, the 

Running Screen 

(Figure 13) 

will appear.  

 

6.  To pause the test, press the 

PAUSE

 button.  

Pressing the 

PAUSE

  button twice will end 

the test and lead the user back to the set up 
screen.  Press the

  CONTINUE

  button to 

continue running the test.  

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

6  Safety Features 

 

The Koehler Water Separability Apparatus is 

equipped with several safety and protection 
features, which are described in the following 
sections. 

 

WARNING:

  The bath is 

NOT

  Explosion 

Proof.  

 

6.1 

Over Temperature Protection 

 

The Koehler Water Separability Apparatus is 

equipped with Over-temperature Protection 
(OTP) circuitry, which prevents the unit from 
exceeding unsafe operating temperatures. If the 
unit cannot maintain the set point temperature 
and begins to decline, the OTP circuitry may 
have been activated. Please follow these steps. 

 

1.  Turn off the unit power switch and disconnect 

the line cord. 

 
2.  Determine the source of the problem and 

correct the situation. 

 
3.  Restart the unit. Monitor the operations to 

ensure that the unit is operating properly. If 
you are still experiencing trouble, please 
contact Koehler technical service for 
assistance. 

 

6.2 

Over Power Protection 

 

The Koehler Water Separability  Apparatus is 

equipped with Over-power Protection circuitry, 
which prevents the unit from unsafe electrical 
conditions. If power to the unit is lost, then turn 
off the main power and turn it back on again. 
The main power switch also functions as a 
circuit breaker. 
 
 
 
 
 
 
 
 
 
 
 
 
 

Summary of Contents for K39400

Page 1: ...venue Bohemia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Analysis...

Page 2: ......

Page 3: ...andards set forth in this certification Specifications ASTM D1401 ASTM D6074 ASTM D6158 ISO 6614 DIN 51599 FTM 791 3201 NF T 60 125 This unit is tested before it leaves the factory to ensure total fun...

Page 4: ......

Page 5: ...tement is given cataloging the items that require separation from the equipment along with basic information on subsequent processing or recycling prior to disposal of the equipment http www dpa syste...

Page 6: ...minimum the following substances preparations and components have to be removed from any separately collected WEEE Mercury containing components such as switches or backlighting lamps compact fluores...

Page 7: ...packing 5 3 3 Setup 6 4 Descriptions 7 4 1 Description of Instrument Controls 7 4 2 Temperature Controller Operation 7 5 Operation 9 5 1 Bath 9 5 2 Power 9 5 4 Test Control Panel Operation 11 6 Safety...

Page 8: ...ducts are backed by our staff of technically knowledgeable trained specialists who are experienced in both petroleum products testing and instrument service to better understand your requirements and...

Page 9: ...Over temperature Protection This unit is equipped with over temperature Protection OTP circuitry to prevent overheating The unit will automatically interrupt power whether equipment malfunction or op...

Page 10: ...red glass front panel jar pose a hazard to personnel Impacts to these during normal operation can shatter the glass components releasing the bath fluid posing thermal burn hazards and slipping hazards...

Page 11: ...g Bath Stainless Steel Bath with clear tempered glass front viewing window for stabilizing the bath temperature to desired setting within 0 05 C 6 Cylinder Ports The test cylinder is placed into the h...

Page 12: ...ture Calculate the difference between the measuring device and the Process value displayed on the controller b Press the program key two times until PCt is displayed in the lower green LED display Pre...

Page 13: ...ium must be changed See Draining instructions in section 7 3 of this document When a decision is made to discard this material dispose of this material and its container to a hazardous of special wast...

Page 14: ...e rail Otherwise shaft may get bent or damaged and will no longer be useful for operation 6 Lift the Stirrer shaft up then move the head assembly left or right until stirrer paddle is over the sample...

Page 15: ...op button may be used to interrupt the 10 second audible alarm 11 At the end of the stirring period please follow the appropriate test method such as those listed in section 1 2 for determining the wa...

Page 16: ...atus is equipped with Over temperature Protection OTP circuitry which prevents the unit from exceeding unsafe operating temperatures If the unit cannot maintain the set point temperature and begins to...

Page 17: ...ior with organic chemicals such as Acetone Toluene Hexane etc For more difficult cleaning of finished surfaces a dilute solution or isopropanol in water may be used It is not recommended that more agg...

Page 18: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 14 8 Wiring Diagrams 8 1 K39400 Unit Wiring...

Page 19: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 15...

Page 20: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 16...

Page 21: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 17 8 2 K39496 Unit Wiring...

Page 22: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 18...

Page 23: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 19...

Page 24: ...eturn of the product to the factory The sole responsibility of Koehler Instrument Company and the purchaser s exclusive remedy for any claim arising out of the purchase of any product is the repair or...

Page 25: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 21 Notes...

Page 26: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 22 Notes...

Reviews: