Koehler K39400 Operation And Instruction Manual Download Page 11

K394XX Water Separability Tester 

Operation and Instruction Manual 

 
 

K394XX-Manual

    

 

-

7-

 

 

4  Descriptions 

 

4.1 

Description of Instrument Controls 

 
 

 

 

Figure 1: Instrument Descriptions 

 
 

1.  Power Switch.

  This switch controls the 

power to the entire unit. When the power 
switch is in the on position, the digital 
temperature controller and test control panel 
are powered on. 
 

2.  Lamp  Switch.

  This switch controls turning 

On and Off the lamp for illuminating the test 
samples. This function available only after 
power switch is turned on.  

 

3.  Temperature Controller.

  The temperature 

controller regulates the bath temperature for 
the test procedure.  Please refer to section 
4.2 for full operational details. 

 

4.  Test Control Panel.  

The test control panel 

regulates the test parameters and stirrer 
paddle activation.  Please refer to section 
4.3 for full operational details. 

 

5.  Heating Bath: 

Stainless  Steel Bath with 

clear tempered glass front viewing window 

for stabilizing  the bath temperature to 
desired setting within ± 0.05°C.  

 

6.  Cylinder Ports: 

The  test  cylinder is placed 

into the heating  bath through one of the 
seven (7) cylinder ports.  
 

7.  Stirrer  Assembly:

  Ensure the stirrer is in 

the upper position and move the entire 
stirrer assembly left or right to align with one 
of the seven test positions. Once aligned 
with desired position, slide stirrer shaft 
downwards into the glass test cylinder. Be 
sure the notches of the stirrer shaft  handle 
align  with the groves on top of the stirrer 
assembly.  

 

8.  Stirrer Shaft Handle. 

For sliding the stirrer 

shaft to the upper standby position or lower 
test position  

 

WARNING:

 

For safety, disconnect the 

power when performing any maintenance and/or 
cleaning.

 

Do 

NOT

  remove cover (bellow) while 

instrument is energized! 
 

4.2 

Temperature Controller Operation 

 

 

 

 

 
 

Figure 2. Temperature Controller 

 

1.  Process  Temperature Display.

  The 

upper red LED  display shows the 
process  temperature as read from the 
RTD probe. 

 

Summary of Contents for K39400

Page 1: ...venue Bohemia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Analysis...

Page 2: ......

Page 3: ...andards set forth in this certification Specifications ASTM D1401 ASTM D6074 ASTM D6158 ISO 6614 DIN 51599 FTM 791 3201 NF T 60 125 This unit is tested before it leaves the factory to ensure total fun...

Page 4: ......

Page 5: ...tement is given cataloging the items that require separation from the equipment along with basic information on subsequent processing or recycling prior to disposal of the equipment http www dpa syste...

Page 6: ...minimum the following substances preparations and components have to be removed from any separately collected WEEE Mercury containing components such as switches or backlighting lamps compact fluores...

Page 7: ...packing 5 3 3 Setup 6 4 Descriptions 7 4 1 Description of Instrument Controls 7 4 2 Temperature Controller Operation 7 5 Operation 9 5 1 Bath 9 5 2 Power 9 5 4 Test Control Panel Operation 11 6 Safety...

Page 8: ...ducts are backed by our staff of technically knowledgeable trained specialists who are experienced in both petroleum products testing and instrument service to better understand your requirements and...

Page 9: ...Over temperature Protection This unit is equipped with over temperature Protection OTP circuitry to prevent overheating The unit will automatically interrupt power whether equipment malfunction or op...

Page 10: ...red glass front panel jar pose a hazard to personnel Impacts to these during normal operation can shatter the glass components releasing the bath fluid posing thermal burn hazards and slipping hazards...

Page 11: ...g Bath Stainless Steel Bath with clear tempered glass front viewing window for stabilizing the bath temperature to desired setting within 0 05 C 6 Cylinder Ports The test cylinder is placed into the h...

Page 12: ...ture Calculate the difference between the measuring device and the Process value displayed on the controller b Press the program key two times until PCt is displayed in the lower green LED display Pre...

Page 13: ...ium must be changed See Draining instructions in section 7 3 of this document When a decision is made to discard this material dispose of this material and its container to a hazardous of special wast...

Page 14: ...e rail Otherwise shaft may get bent or damaged and will no longer be useful for operation 6 Lift the Stirrer shaft up then move the head assembly left or right until stirrer paddle is over the sample...

Page 15: ...op button may be used to interrupt the 10 second audible alarm 11 At the end of the stirring period please follow the appropriate test method such as those listed in section 1 2 for determining the wa...

Page 16: ...atus is equipped with Over temperature Protection OTP circuitry which prevents the unit from exceeding unsafe operating temperatures If the unit cannot maintain the set point temperature and begins to...

Page 17: ...ior with organic chemicals such as Acetone Toluene Hexane etc For more difficult cleaning of finished surfaces a dilute solution or isopropanol in water may be used It is not recommended that more agg...

Page 18: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 14 8 Wiring Diagrams 8 1 K39400 Unit Wiring...

Page 19: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 15...

Page 20: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 16...

Page 21: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 17 8 2 K39496 Unit Wiring...

Page 22: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 18...

Page 23: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 19...

Page 24: ...eturn of the product to the factory The sole responsibility of Koehler Instrument Company and the purchaser s exclusive remedy for any claim arising out of the purchase of any product is the repair or...

Page 25: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 21 Notes...

Page 26: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 22 Notes...

Reviews: