Koehler ADA5000 Operation And Instruction Manual Download Page 5

ADA5000 Automatic Distillation Analyzer 

Operation and Instruction Manual 

ADA5000-Manual 

2 -

 

 

1.3 Instrument Specifications 

 

Models: K45603, 

K45604 

K45703-TS, K45704-TS 
 

Electrical Requirements:  120VAC ± 5%, 50/60Hz 

230VAC ± 5%, 50/60Hz 
 

Distillation Range: 

0 - 450°C (±0.1°C) 
 

Condenser: 

-5 to 60°C (±0.1°C) 
 

Receiver Chamber: 

0 to 60 C (±0.1°C) 
 

Distillation Rate: 

2-15 mL/min in 0.01mL 
increments 
 

Receiver Volume: 

0 to 100 mL 
 

Barometric Pressure: 

550 to 900 mm Hg (±1 mm Hg) 
 

Dimensions (l x w x h) 

21 x 21.5 x 27 in (53.3 x 54.6 x 
70.5 cm) 
 

Net Weight: 

200 lbs (91kg) 
 

Operational Conditions:  0-45°C (32 to 113°F) 

 

1.4 Software Specifications 

 

PC System 
Requirements: 

Intel Pentium III Processor or 
similar (minimum) 
 

Operating System: 

Windows 98/XP 
 

Memory 
Requirements: 

128 Mb RAM 
4Gb hard drive 
 

Communication Ports:   RS-232 (serial), USB, RJ45 

(Ethernet) 

 

2  Safety Information and Warnings 

 

Safety Considerations. 

The use of this 

equipment may involve hazardous materials and 
operations. This manual does not purport to 
address all of the safety problems associated 
with the use of this equipment. It is the 
responsibility of any user of this equipment to 
investigate, research, and establish appropriate 
safety and health practices and determine the 
applicability of regulatory limitations prior to use. 
 

Equipment Modifications and Replacement 

Parts. 

Any modification or alteration of this 

equipment from that of factory specifications is 
not recommended, voids the manufacturer 
warranty, product safety, performance 
specifications, and/or certifications whether 

specified or implied, and may result in personal 
injury and/or property loss. Replacement parts 
must be O.E.M. exact replacement equipment. 
 

Internal Access.

 If internal access into the 

instrument is necessary, the analyzer must be 
disconnected from the main power supply 
otherwise electrical hazards may be 
experienced. Be extremely careful because of 
electrical shock and danger of personal injury. 
 

Unit Design. 

This equipment is specifically 

designed for use in accordance with the 
applicable standard test methods listed in 
section 1.2 of the operator’s manual. The use of 
this equipment in accordance with any other test 
procedures, or for any other purpose, is not 
recommended and may be extremely 
hazardous.  
 

Fire Safety. 

This equipment is designed with 

an automatic fire extinguishing system. The 
system must be checked periodically for proper 
operation. 
 

Troubleshooting. 

 In the case of 

troubleshooting the instrument, it is important to 
be careful with the electronic boards. They are 
static sensitive and it is absolutely necessary 
that the operator has to be grounded before 
touching the boards. 
 

Overtemperature Protection.

 This unit is 

equipped with Overtemperature Protection 
(OTP) circuitry to prevent overheating. The unit 
will automatically interrupt power whether 
equipment malfunction or operator error causes 
the temperature to exceed the maximum 
recommended temperature range. The power 
can only then be restored by identifying and 
correcting the problem, allowing the unit to 
return to normal operating temperatures, and 
resetting the power to the unit. 
 

Heater Element. 

The heater element (heating 

coil) of the distillation unit, installed in the 
heating area within the flask support has to be 
inspected at least every 3 months. If any short 
circuits of the windings are observed, it is 
necessary to use a new heater element. It is 
important to do this exchange very carefully. The 
windings must not have a short circuit and need 
to be adjusted carefully before operation. Do 
NOT touch the heater element after a distillation 
run. The element could still be extremely hot and 
one could experience personal injuries. 
 

Summary of Contents for ADA5000

Page 1: ...venue Bohemia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Analysis...

Page 2: ......

Page 3: ...3 1 Packing List 3 3 2 Unpacking 3 3 3 Instrument Installation 3 3 4 Software Installation 3 4 Operation 7 4 1 Description of Instrument Controls 7 4 2 Instrument Operation 7 4 3 Communication Manage...

Page 4: ...o are experienced in both petroleum products testing and instrument service to better understand your requirements and provide you with the best solutions You can depend on Koehler for a full range of...

Page 5: ...e analyzer must be disconnected from the main power supply otherwise electrical hazards may be experienced Be extremely careful because of electrical shock and danger of personal injury Unit Design Th...

Page 6: ...different models are K45603 K45703 TS 120VAC 5 50 60Hz 20A K45604 K45704 TS 230VAC 5 50 60Hz 10A Nitrogen Nitrogen must be available in order for the automatic fire extinguishing system to work Conne...

Page 7: ...nt by clicking on the Yes button Confirm that the computer meets these minimum requirements and click on the Next button Enter the name and company information of the user To utilize the software as a...

Page 8: ...ext button to continue Verify that all each box of the four components are checked and click on the Next button Select or create a folder the add program icons to Click Next to continue Review the pro...

Page 9: ...arious paths please consult with Koehler technical service To change the units for temperature 1 and pressure 2 select the desired units by clicking on the corresponding radio buttons Registration 3 F...

Page 10: ...amber 4 2 Instrument Operation 1 Power Turn on the main power switch to the unit Turn on the power switches on the PC and printer if applicable Note the internal computer on the K457XX TS models has a...

Page 11: ...correct position Place the silicone stopper over the side arm of the distillation flask and then place it into the condenser tube 13 Move the heater carefully under the flask 14 Place the dropping pl...

Page 12: ...e the status of an analyzer from the PC select the View option and then Unit Data Select unit 1 Select a connected analyzer to see all the important values of this analyzer Analog 2 Temperature shown...

Page 13: ...sible to change the path of the database program form and results form For a normal installation you can use this standard directory If a PC is connected to a network it s possible to have the databas...

Page 14: ...ults Form Select this item to change a result form Be careful when changing a form A backup of these forms can be found on the installation CD in directory Backup Results Comparison To compare up to 4...

Page 15: ...ram Unit Select analyzer Load Select item to load a new program from the database Attention Program will be sent automatically to selected analyzer and overwrite an existing previous run If desired st...

Page 16: ...Koehler Comment Sample Program Method ASTM D86 Distillation Group SG4 Condenser temp 40 0 C Receiver temp 30 0 C Expected volume 98 0 mL Preheat 1 29 1 Preheat time 3 0 min Preheat 2 28 2 Extended Fi...

Page 17: ...finished min 30 mL are necessary microprocessor system of distillation analyzer will calculate optimized value According to the distillation group a heating rate will be calculated to get a time for...

Page 18: ...fix temperatures are stored in a memo field of the database and are online available Program Extended Fix Temps Final Results All important values are shown online here That means if the IBP was detec...

Page 19: ...since start is shown Cool Time Timer for cooling device Condenser Shows actual temperature of condenser tube Volume Shows actual volume of selected analyzer Receiver Shows actual temperature of recei...

Page 20: ...wing sections 5 1 Flask Heater Compartment and Auto Fire Extinguisher The design of the flask heater compartment makes this instrument one of the safest automatic distillation units available The phys...

Page 21: ...in personal injury or death If you have any questions about maintaining your equipment please do not hesitate to contact the Koehler technical service department 6 1 Routine Maintenance The Koehler A...

Page 22: ...s been properly installed operated and maintained Koehler Instrument Company must be advised in writing of the malfunction and authorize the return of the product to the factory The sole responsibilit...

Page 23: ...ADA5000 Automatic Distillation Analyzer Operation and Instruction Manual ADA5000 Manual 20 Notes...