16 -

 

 

 
Unit: 

Select analyzer 

 
Start/Stop button: 

Click button, to start or to 

stop a distillation run of selected analyzer. 
Status of analyzer will be shown. 

 
Temperature

: Temperatures in °C or °F are 

possible. Please note, that program works 
internally with °C and all temperatures stored in 
the data base are also in °C. If a distillation run 
is to be recalled, it is shown in the units that it 
was run. A recalled distillation run made in °F 
will shown in °F. To export values out of the 
ACCESS

 

database by an installed full version of 

ACCESS, it must manually calculated to °F if 
desired. 
 

Pressure

: Pressure in hPa and mmHg can be 

selected. 
 

Flask: 

Actual flask temperature is shown 

according selected unit. 

 
Baro: 

Actual pressure is shown according 

selected unit. 

 
Dist Time: 

Complete distillation time since start 

is shown. 

 
Cool Time: 

Timer for cooling device.  

 
Condenser: 

Shows actual temperature of 

condenser tube 

 
Volume: 

Shows actual volume of selected 

analyzer 

 
Receiver: 

Shows actual temperature of receiver 

compartment 

 

Dist. Rate: 

Shows actual dist. rate 

 
Heat: 

Shows actual heat capacity. 

 

4.6 Diskstat.exe 

 
A graphical overview of all connected 

analyzers is shown. If “AutoSave“ is selected, 
then it is necessary to start this program. At 
start-up of the program, it’s possible that an 
error-message appears with “PT100 defective“. 
Select OK (three times) to run the program. 
 

 

 

4.7 Config.exe 

 

It is possible to store the database and forms 

in any chosen directory. In a normal case use 
the local directory. Koehler software can also be 
used in a network system. 

DataPath - Program-DB - Result-DB - 
ProgramForm - ResultsForm 

 
Temp Unit: 

Select unit: °C °F 

 
Baro Unit:  

Select unit hPa or mmHg 

 
Name: 

Input your name 

 

Company:

 Input company name exactly as is 

shown on the registration form 
 

Serial No:

 Input serial number exactly as is 

shown on the registration form. 
 
Compare “Install of software“ for details. 
 

Summary of Contents for ADA5000

Page 1: ...venue Bohemia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Analysis...

Page 2: ......

Page 3: ...3 1 Packing List 3 3 2 Unpacking 3 3 3 Instrument Installation 3 3 4 Software Installation 3 4 Operation 7 4 1 Description of Instrument Controls 7 4 2 Instrument Operation 7 4 3 Communication Manage...

Page 4: ...o are experienced in both petroleum products testing and instrument service to better understand your requirements and provide you with the best solutions You can depend on Koehler for a full range of...

Page 5: ...e analyzer must be disconnected from the main power supply otherwise electrical hazards may be experienced Be extremely careful because of electrical shock and danger of personal injury Unit Design Th...

Page 6: ...different models are K45603 K45703 TS 120VAC 5 50 60Hz 20A K45604 K45704 TS 230VAC 5 50 60Hz 10A Nitrogen Nitrogen must be available in order for the automatic fire extinguishing system to work Conne...

Page 7: ...nt by clicking on the Yes button Confirm that the computer meets these minimum requirements and click on the Next button Enter the name and company information of the user To utilize the software as a...

Page 8: ...ext button to continue Verify that all each box of the four components are checked and click on the Next button Select or create a folder the add program icons to Click Next to continue Review the pro...

Page 9: ...arious paths please consult with Koehler technical service To change the units for temperature 1 and pressure 2 select the desired units by clicking on the corresponding radio buttons Registration 3 F...

Page 10: ...amber 4 2 Instrument Operation 1 Power Turn on the main power switch to the unit Turn on the power switches on the PC and printer if applicable Note the internal computer on the K457XX TS models has a...

Page 11: ...correct position Place the silicone stopper over the side arm of the distillation flask and then place it into the condenser tube 13 Move the heater carefully under the flask 14 Place the dropping pl...

Page 12: ...e the status of an analyzer from the PC select the View option and then Unit Data Select unit 1 Select a connected analyzer to see all the important values of this analyzer Analog 2 Temperature shown...

Page 13: ...sible to change the path of the database program form and results form For a normal installation you can use this standard directory If a PC is connected to a network it s possible to have the databas...

Page 14: ...ults Form Select this item to change a result form Be careful when changing a form A backup of these forms can be found on the installation CD in directory Backup Results Comparison To compare up to 4...

Page 15: ...ram Unit Select analyzer Load Select item to load a new program from the database Attention Program will be sent automatically to selected analyzer and overwrite an existing previous run If desired st...

Page 16: ...Koehler Comment Sample Program Method ASTM D86 Distillation Group SG4 Condenser temp 40 0 C Receiver temp 30 0 C Expected volume 98 0 mL Preheat 1 29 1 Preheat time 3 0 min Preheat 2 28 2 Extended Fi...

Page 17: ...finished min 30 mL are necessary microprocessor system of distillation analyzer will calculate optimized value According to the distillation group a heating rate will be calculated to get a time for...

Page 18: ...fix temperatures are stored in a memo field of the database and are online available Program Extended Fix Temps Final Results All important values are shown online here That means if the IBP was detec...

Page 19: ...since start is shown Cool Time Timer for cooling device Condenser Shows actual temperature of condenser tube Volume Shows actual volume of selected analyzer Receiver Shows actual temperature of recei...

Page 20: ...wing sections 5 1 Flask Heater Compartment and Auto Fire Extinguisher The design of the flask heater compartment makes this instrument one of the safest automatic distillation units available The phys...

Page 21: ...in personal injury or death If you have any questions about maintaining your equipment please do not hesitate to contact the Koehler technical service department 6 1 Routine Maintenance The Koehler A...

Page 22: ...s been properly installed operated and maintained Koehler Instrument Company must be advised in writing of the malfunction and authorize the return of the product to the factory The sole responsibilit...

Page 23: ...ADA5000 Automatic Distillation Analyzer Operation and Instruction Manual ADA5000 Manual 20 Notes...

Reviews: