6 Menu overview
32
Parameter
Description
QAFAIL
Output behavior as per NE43 in the event of a fault (function only available when
current output has been selected under QATYP).
●
3.5 mA = Analog current output is set to 3.5 mA in the event of a fault
●
21.5 mA = Analog current output is set to 21.5 mA in the event of a fault
SimCur
See Chapter 5.3 "Testing the configuration"
SimVol
See Chapter 5.3 "Testing the configuration"
DspVal
Display settings
●
Distan = The display shows the distance in mm in relation to the end of the
probe.
●
QaPerc = The display shows the fill level in % in relation to the QA analog
output with the corresponding QAHIGH and QALOW thresholds.
●
QaBarG = The display shows a bar graph in relation to the QA analog output
with the corresponding QAHIGH and QALOW thresholds.
●
QaSign = The display shows the current QA output value in mA or V.
●
QxSign = The display shows the output states.
Filter
See Chapter 5.2 "Filtering measured values"
SimLev
See Chapter 5.3 "Testing the configuration"
RstFac
Resetting of the set parameters back to the factory settings