12. Input adjustment
The zero point and slope of each measuring inputs can be adjusted with the
parameters offset (“OFFS”) and scale (“SCAL”).
A reasonable adjustment presumes reliable references (e.g. ice water, controlled
precision water bath, etc.). If the inputs are adjusted (i.e. offset and scale are
different from default settings) the device will shortly display “Corr” after turned
on.
Default setting for offset and scale are ‘off’ = 0.0, i.e. inputs are not changed.
Zero point correction:
Displayed value = measured value – OFFS
Zero point and slope correction:
Displayed value = (measured value – OFFS) * (1 + SCAL / 100)
Displayed value °F = (meas. value °F - 32°F - OFFS) • (1 + SCAL / 100)