HND-R106
page 18
HND-R106 K09/0422
10. Configuration
Some menu points depend on current device settings.
To change device settings, press
“Menu”
for 2 seconds. This will activate the
configuration menu (main display: “SEt”). Pressing
“Menu”
changes between
the menus points, pressing
jumps to the referring parameters, which can be
selected with key
.
The parameter can be changed with
or
. Pressing
“Menu”
again jumps
back to the main configuration menu and saves the settings. Pressing “Quit”
finishes the configuration and returns back to standard measuring operation.
Pressing “menu” and “store” at the same time for more than 2 seconds will
reset the device to factory defaults.
If no key is pressed for more than 2 minutes the configuration will be aborted. All
changes will not be saved!
Menu
Parameter Value
Description
or
.
Set Configuration: General configuration
Input: Selection of measured values
Arrow
„rh“
rH value measurement
Arrow
„mV“
mV value measurement (REDOX or ORP)
Arrow
„mV
H
“
mV value measurement referring to
standard hydrogen system
Arrow
„pH“
pH value measurement
Resolution pH: resolution of pH display
0.1..0.01
tenth pH..hundredth pH
Calibration: Select number of calibration points
1-Pt
1-point (only offset calibration, slope = -59.2
mV/pH)
2-Pt
2-point (n another one)
3-Pt
3-point (n one acid + one alkaline
buffer)