Kobold HND-F215 Operating Instructions Manual Download Page 13

 

 

HND-F215 

HND-F215 K07/0522 

 

page 13 

Flow  measuring  probes  are  'free-jet  calibrated',  i.e.  the  diameter  of  the  flow 
channel has to be 5 times bigger than the diameter of the flow measuring head (= 
approx. 5 cm, otherwise measuring errors up to 40 %). 
When  evaluating  the  measuring  results  please  also  note  that  in  a  channel  the 
flow  speed  is  usually  higher  in  the  middle  of  the  channel  than  at  its  edges. 
Therefore, use appropriate tables to calculate air flow by means of flow speed. 
 

Averaging for Flow Measurements: 

When taking flow measurements fluctuations tend to be quite high. To be able to 
display a stable measuring value two averaging functions have been integrated in 
the instrument. 
 

Continuous Averaging 

The  average  value  displayed  has  been  calculated  from  the  past  few 
measurements conducted during the averaging time set After the instrument has 
been switched on the time remaining till expiration of the averaging time will be 
displayed at the bottom line of the display. The min. and max. values memorized 
refer to the minimum and/or maximum average value displayed. 
 

Average Hold 

As  soon  as  the  HND-F215  instrument  has  been  switched  on  the  device  starts 
calculating  the  average  flow  value  during  the  averaging  time.  During  measuring 
the 

current measuring value

 will be shown in the top line of the display while the 

bottom  line  shows  the  remaining  measuring  time.  As  soon  as  measurements 
have  been  completed  the 

average  value

  will  be  displayed  and  the  device  will 

switch  to  the  HOLD  mode.  The  min.  and  max.  values  memorized  refer  to  the 
minimum and/or maximum measuring value established during averaging. 
To start a new measuring series press the key "Store" (key 6). 
 

Additional Measurements with any NiCr-Ni-Temperature Probe at 
T2: 

Use temperature channel T2 to take measurements of medium temperature, for 
example. The value shown is not an average value. 
 
 

7.8  Notes for Special Functions 

Zero displacement ('Offset') 

A zero displacement can be carried out for each of the two temperature channels 
T1 (HND-FF31 only) and T2: 
 

displayed temperature = measured temperature - Offset 

 
Standard  setting:  'off'  =  0.0°,  i.e.  no  zero  displacement  will  be  carried  out.  The 
zero  displacement  is  mainly  used  to  compensate  for  sensor  deviations.  Unless 
'off'  is  set,  this  value  will  be  displayed  shortly  after  the  device  is  switched  on; 
during operation it will be identified by means of the offset arrow in the display. 
 

Summary of Contents for HND-F215

Page 1: ...Operating Instructions for Manual Humidity Precision Measuring Unit Model HND F215...

Page 2: ...tes for Special Functions 13 7 9 The Serial Interface 17 7 10 How to Calibrate Meas of Rel Humidity Using HND FF31 18 7 11 System and Error Messages 20 7 12 System and Error Messages 21 7 13 Error and...

Page 3: ...used maintained and serviced by persons familiar with these operating instructions and in accordance with local regulations applying to Health Safety and prevention of accidents When used in machines...

Page 4: ...ng voltage has to be 10 5 to 12 VDC Do not apply over voltage Cheap 12 V power supply devices often have excessive no load voltage We therefore recommend using regulated voltage power supply devices T...

Page 5: ...y has to be designed most carefully Internal connection in third party devices e g connection GND and earth may result in not permissible voltages impairing or destroying the device or another device...

Page 6: ...F32 flow speed water 0 05 5m s 7 3 Displays Depending on the measuring probes sensors connected the following measuring results can be displayed HND FF31 Main display r H relative atmospheric humidity...

Page 7: ...arrow indicates that correction factor is activated Logger arrow indicates that the logger function is activated Alarm arrow indicates an alarm Messages at device startup The device will show some mes...

Page 8: ...g probe only press for 2 sec humidity calibration will be started press for more than 10 sec reset of humidity calibration to factory calibration Set Menu press Set shortly display changes between T1...

Page 9: ...s marked by 1 will only be displayed if no data is stored in the logger AVG Selection of Averaging Proceedings for Flow Measurement 1 only HND FF32 33 Cont continuous averaging the average value calcu...

Page 10: ...the base addresses accordingly Set Alarm Alarm Settings Settings for the alarm function Please note the points marked by 2 will only be displayed if the alarm functions on or no So have been selected...

Page 11: ...The scope of supply includes one sensor to measure relative atmospheric humidity and another one to measure the ambient temperature T1 rel humidity r H Relative humidity measured in the tip of the pro...

Page 12: ...T1 T2 and relative atmospheric humidity The combination sensor is used to measure the ambient air whose condition is issued to calculate the dew point Td The surface sensor is used to measure surface...

Page 13: ...s the HND F215 instrument has been switched on the device starts calculating the average flow value during the averaging time During measuring the current measuring value will be shown in the top line...

Page 14: ...ls 2 and 3 will have the following addresses Example base address 21 channel 1 21 channel 2 22 channel 3 23 Alarm 3 alarm settings are available off off on with horn sound on on no horn sound no So De...

Page 15: ...d time date Press Store key to store current values St XX will be displayed for a short time XX representing the number of the data set 1 99 If the logger memory is full a warning will appear on the d...

Page 16: ...g value Stop logger recording Press Store key key 6 for a short time to stop recording You will then be asked to acknowledge again recording to be stopped recording to be continued Use the keys key 2...

Page 17: ...Read offset correction x 218 Read corr factor 1000 1200 3 3 3 3 3 3 224 Read logger data cyclic logger 4 225 Read logger cycle 5 226 Set logger cycle 6 227 Start logger recording 7 228 Read count of l...

Page 18: ...libration equipment is quite often highly dependent on temperature This dependence is automatically compensated for when calibrating with the integrated calibration equipment and automatic detection I...

Page 19: ...detected humidity value measured may deviate from value set by manufacturer by approx 10 In case of manual input enter value here As soon as the display stops blinking and changing between values a st...

Page 20: ...OC xx xx identified sensor HND FF31 FF32 or FF33 temperature offset of the HND FF31 sensor only with HND FF31 and adj offset value off flow averaging procedure only with HND FF32 FF33 AVG Hold or AVG...

Page 21: ...mains op power supply defective or wrong voltage polarity check replace power supply System error disconnect battery or power supply wait for a short time re connect device defective return to manufa...

Page 22: ...the battery has been used up and needs to be replaced The device will however operate correctly for a certain time If bAt is shown in the upper display the voltage is too low to operate the device th...

Page 23: ...onal Power supply 9 V monobloc battery included in the scope of delivery external 10 5 12 VDC via jack Material housing made of impact resistant ABS plastic Protection IP 65 front Dimensions 142 x 71...

Page 24: ...sor head and tube 40 60 C short term 120 C Dimensions probe 14 x 119 mm handle 19 x 135 mm 1 m cable HND FF32 Flow meas probe with spring head calibrated and completely replaceable Measuring range 0 0...

Page 25: ...rter on USB interface HND Z034 Windows software for setting data read out and printing of the data of housings of the HND series with logger function HND Z052 100 ml conductivity paste for surface mea...

Page 26: ...low EN 61326 1 2013 Electrical equipment for measurement control and laboratory use EMC requirements Part 1 General requirements EN 50581 2012 Technical documentation for the assessment of electrical...

Page 27: ...n Hard Acronychia laevis h 28 5 66 Ayan Distemonanthus benthamianus h 285 7 54 Balau Shorea laevis h 31 4 54 Balau red Shorea guiso h 32 4 68 Balsa Ochroma pyramidale h 33 4 91 Basralocus Angelique Di...

Page 28: ...Cypress Northern Callitris intratropica h 115 6 78 Cypress Rottnest Island Callitris preisii h 116 7 80 Cypress White Callitris glaucophylla h 117 6 86 Dakua Salusalu Fiji Decussocarpus vitiensis h 1...

Page 29: ...leri h 243 8 73 Mahogany New Guinea Dysoxylum spp h 241 6 74 Mahogany Red Eucalyptus botryoides h 244 7 91 Mahogany Rose Dysoxylum fraseranum h 245 7 65 Mahogany Southern Eucalyptus botryoides h 246 5...

Page 30: ...e Celery Top Phyllocladus aspenifolius h 329 7 71 Pine Hoop Araucaria cunninghamii h 330 7 79 Pine Huon Dacrydium franklinii h 331 8 70 Pine King William Athrotaxis selaginoides h 332 7 67 Pine Klinki...

Page 31: ...atopetalum succirubrum h 408 7 63 Tallowwood Eucalyptus microcorsis h 409 4 92 Tatajuba Bagassa guianesis h 30 7 44 Taun Maleisien Pometia pinnata h 195 0 105 Taun New Guinea Pometia pinnata h 196 6 1...

Page 32: ...7 yes Plaster synthetic b 12 18 2 60 8 yes On wall plaster b 20 0 0 38 8 no Lime mortar KM 1 3 b 14 0 4 40 4 yes Lime sand bricks 14 DF 200 gross density 1 9 b 28 0 1 12 5 yes Limestone b 15 0 4 29 5...

Reviews: