Supplement manual Foundation Fieldbus 

 

 

 

ft3. 

 
Sensor constant

: the sensor constant of the transmitter can be read out and can be changed if necessary (e. g. at a 

re-calibration of the sensor) 

 
Fixed density:

 If the measurement of the density is not possible you can preset a fixed factor for calculation mass 

flow into volume flow. Dimension g/l is always used as fixed density factor. 

 
Density measurement:

 parameter to switch on or off the density measurement or to use the factor of the fixed 

density. If device is shipped with a factory density calibration, the parameter is set to on.

 

If device is shipped 

without a density calibration, it is not possible to switch on the measurement.  Only the fixed density can be used. 
 

 Empty pipe limit:

 below this limit it is assumed that there is no medium in the loops. The status of the output 

will be set to “uncertain“. This parameter is given in g/l. 

 
Phase:

 display of the actual phase shifting (to be used for diagnosis) 

 

 Frequency:

 display of the actual frequency (to be used for diagnosis). 

 

 Sensor amplitude:

 display of the actual sensor amplitude (to be used for diagnosis).  

 

Exciter current

: display of the actual exciter current (to be used for diagnosis). 

 

Max. electronic temperature:

 display of the highest measured temperature inside the transmitter.  

 

Max. measured medium temperature:

  display of the highest measured temperature of the medium. 

  

Totalizer forward flow:

 display of forward flow totalizer. 

 

Totalizer reverse flow:

 display of reverse flow totalizer. 

 

Reset totalizer:

 Set the parameter to “reset” and write into device. Both totalizer (forward and reverse) are reset 

to Zero. 
 

Calibrate zero.

 Set the parameter to “calibrate“ and write into device. Phase shifting offset will be compensated 

by an automatically procedure (zero point adjustment). The procedure lasts app. 30seconds.  
 

System error:

  system errors are serious malfunctions of the device and should never occur during normal 

processing. Please note the error number before clearing it. Please call Heinrichs Messtechnik Service and send 
this error number. The coding of these errors can be derived from the standard manual. 
 

Reset system errors:

 Reset any system errors. 

 

Status message:

 bit coded status of the transmitter. For diagnostic purpose only. 

 

Status:

 in case of a malfunction you will get information about the reason of this malfunction. 

 
 
 
 
 

Summary of Contents for Heinrichs UMC3

Page 1: ...nt manual Foundation Fieldbus Rev No 20 01 Page 1 of 10 Operating manual for the electronic transmitter UMC3 FF with Foundation Fieldbus interface UMC3 Supplement to the TM UMC3 TMU UMC3 TMR UMC3 and...

Page 2: ...VALUES _________________________________ 4 3 1 Supply voltage 4 3 2 Basic current 4 3 3 Baud rate 4 3 4 Electromagnetic compatibility EMC 4 3 5 Safety data 4 4 ELECTRICAL CONNECTION __________________...

Page 3: ...media 2 Introduction The Foundation Fieldbus interface of the UMC3 FF offers the possibility to read out the measured values Parameters can be read and changed if needed The interface fulfills the re...

Page 4: ...ting manual UMC3 3 5 Safety data Refer to additional operating manual for explosion proof flow meters FF communications circuit terminals 55 and 56 For connecting a certified FF circuit in accordance...

Page 5: ...inrichs Messtechnik GmbH Supplement manual Foundation Fieldbus Page 5 of 10 4 Electrical connection 4 1 Shielding The shield of the lines at terminals 55 and 56 must be grounded at least at the PLC si...

Page 6: ...or specific parameters as well as the measured values flow density temperature The Analog Input function block AI contains parameters forming the output value limits range low flow cutoff damping valu...

Page 7: ...surement has been calibrated and is switched on or a fixed density is preset Otherwise the output of the AI will be marked as bad Flow unit can be chosen from the following units Mass flow kg s kg min...

Page 8: ...play of the actual sensor amplitude to be used for diagnosis Exciter current display of the actual exciter current to be used for diagnosis Max electronic temperature display of the highest measured t...

Page 9: ...al to the unit selected in the transducer block If output of AI shall be identically to its input scaling 1 following parameters have to be set to OUT SCALE equal to XD SCALE Any unit of OUT value can...

Page 10: ...transducer block will be set Attention XD_SCALE unit in the AI must be set equal via Fieldbus settings Some selectable units of transmitter UMC3 are not available at Foundation Field bus If such a uni...

Reviews: