Kobold Heinrichs UMC3 Operating Manual Download Page 7

 
 
 

Heinrichs Messtechnik GmbH

 

 

Supplement manual Foundation Fieldbus 

Page

 

7 of 10 

 

5.1 

Device description files 

To declare the function of UMC3-FF to a PLC the device description files are needed. These files can 
be downloaded from the homepage of the Fieldbus Foundation: 

http://www.fieldbus.org/index.php?option=com_mtree&task=search&Itemid=324&ffbstatus=Registered
&ffbmfg=HeMessGmbH&ffbcategory=

 

 

010101.cff 

 

0101.sym 

 

0101.ffo 

5.2 

Device specific parameters of the Transducer block 

SERIAL_NUMBER

: device factory serial number

 

 
Flow unit: 

it can be chosen between units for mass flow or volume flow. Depending on this selection mass flow 

or  volume  flow  will  be  forwarded  to  the  AI.  The  output  of  volume  flow  is  only  possible  if  the  density 
measurement has been calibrated and is switched on or a fixed density is preset. Otherwise the output of the AI 
will be marked as “bad“. Flow unit can be chosen from the following units: 

Mass flow

:  

 

kg/s,  kg/min, kg/h, kg/d 

 

t/min, t/h, t/d, 

 

g/s, g/min, g/h, 

 

 lbs/s, lbs/min, lbs/h, lbs/d, 

 

ston/min, ston/h, ston/d, 

 

lton/h, lton/d,  

Volume flow:  

 

m3/s, m3/min, m3/h, m3/d,  

 

l/s, l/min, l/h,  

 

USG/min, USG/h, MG/d, 

 

UKG/s, UKG/min, UKG/h,  

 

ft3/s, ft3/min, ft3/h, ft3/d,  

 

USB/s, USB/min, USB/h, USB/d 

 
Density unit:

  can be chosen from the following units: 

 

kg/m3 

 

kg/l,  

 

g/l,  

 

g/cm3,  

 

lb/ft3,  

 

lb/USG,  

Temperature unit: 

can be chosen from the following units

 

°C,  

 

F,  

 

K. 

 
Totalizer  unit:

  depending  on  this  unit  it  is  accumulated  mass  or  volume  flow.  The  accumulation  of  volume is 

only  possible  if  the  density  measurement  has  been  calibrated  and  is  switched  on  or  a  fixed  density  is  preset. 
Otherwise the output of the AI will be marked as “bad“.  
The unit can be chosen from the units following:  

 

g, kg, t,  

 

l, m3,  

 

USG,  

 

UKG,  

 

USB,  

 

lbs,  

 

shton,  

 

lton,  

Summary of Contents for Heinrichs UMC3

Page 1: ...nt manual Foundation Fieldbus Rev No 20 01 Page 1 of 10 Operating manual for the electronic transmitter UMC3 FF with Foundation Fieldbus interface UMC3 Supplement to the TM UMC3 TMU UMC3 TMR UMC3 and...

Page 2: ...VALUES _________________________________ 4 3 1 Supply voltage 4 3 2 Basic current 4 3 3 Baud rate 4 3 4 Electromagnetic compatibility EMC 4 3 5 Safety data 4 4 ELECTRICAL CONNECTION __________________...

Page 3: ...media 2 Introduction The Foundation Fieldbus interface of the UMC3 FF offers the possibility to read out the measured values Parameters can be read and changed if needed The interface fulfills the re...

Page 4: ...ting manual UMC3 3 5 Safety data Refer to additional operating manual for explosion proof flow meters FF communications circuit terminals 55 and 56 For connecting a certified FF circuit in accordance...

Page 5: ...inrichs Messtechnik GmbH Supplement manual Foundation Fieldbus Page 5 of 10 4 Electrical connection 4 1 Shielding The shield of the lines at terminals 55 and 56 must be grounded at least at the PLC si...

Page 6: ...or specific parameters as well as the measured values flow density temperature The Analog Input function block AI contains parameters forming the output value limits range low flow cutoff damping valu...

Page 7: ...surement has been calibrated and is switched on or a fixed density is preset Otherwise the output of the AI will be marked as bad Flow unit can be chosen from the following units Mass flow kg s kg min...

Page 8: ...play of the actual sensor amplitude to be used for diagnosis Exciter current display of the actual exciter current to be used for diagnosis Max electronic temperature display of the highest measured t...

Page 9: ...al to the unit selected in the transducer block If output of AI shall be identically to its input scaling 1 following parameters have to be set to OUT SCALE equal to XD SCALE Any unit of OUT value can...

Page 10: ...transducer block will be set Attention XD_SCALE unit in the AI must be set equal via Fieldbus settings Some selectable units of transmitter UMC3 are not available at Foundation Field bus If such a uni...

Reviews: