Heinrichs Messtechnik 

TSK1 Installation and Operating Instructions 

 

Page 4 of 27  

 

Introduction 

 
These Installation and Operating Instructions serve as a tool 
for the correct installation, operation and maintenance of the 
device.  They are a supplement to the TSK Device 
Description.  Read the manual carefully before the device is 
installed and put into use.  It does not include special 
versions or applications. 
 
All devices are thoroughly checked for order compliance and 
operability before delivery.  Upon receipt, please conduct a 
visual inspection of possible damage that may be identified 
as having occurred during shipment.  If you discover any 
defect, please contact our head office in Cologne or the local 
sales office responsible for your area (see the telephone 
directory at the end of this manual or on our Web site).  
Apart from a description of the error, we will need the 
equipment type and serial number of the delivery.  
 
Heinrichs Messtechnik shall not furnish guarantee for any 
repair work done without prior notice.  Unless otherwise 
agreed on, the rejected parts must be made available to us 
in case a claim is made. 
 

Identification 

1.1 

Supplier/manufacturer 

Heinrichs Messtechnik GmbH 
Robert-Perthel-Str. 9   D-50739 Köln 
Phone  +49 (221) 49708 - 0 
Fax      +49 (221) 49708 - 178 
Internet: 

http://www.heinrichs.eu/

 

E-mail

mailto:[email protected]

 

1.2 

Product type 

Paddle-type flowmeter in all-metal design  

1.3 

Product name 

TSK1 with subgroups 
TSK1-C (stainless steel) 
TSK1-S (stainless steel) 
TSK1-K (polypropylene/stainless steel) 
TSK1-P (PTFE/Hastelloy C) 

1.4 

Issue date  

20/10/2008 

1.5 

Version no. 

3.0 
File: TSK1_BA_03_eng 
 

Applications 

The TSK meter is suitable for flow measurement of liquid 
products in pipes.  

The special advantage is that it can be 

used for all directions of flow.  

It shows the current flow 

rate in volume or mass per unit in time. 
 
Applications:  flow measurement, dosing, monitoring, 
adjusting and control of liquid and gaseous products. 
The meter’s design makes it ideal for processes under 
difficult and adverse operating conditions. 
 
The devices are available with additional electrical 
equipment for process monitoring and control. 
 

Operational mode and system design  

3.1 

Measuring principle 

If a medium flows with sufficient velocity through the 
horizontally or vertically arranged measuring chamber (1) if 
the TSK fitting, the paddle (2) will swivel around the axle until 
the force of the medium and the opposing force of the 
paddle surface plus the spring tension establish equilibrium.    
The angular position or the position of equilibrium ( ) of the  
paddle (2) in the measuring chamber (1) is the measure for 
the flow. The encapsulated ring-type permanent magnet (4) 
at the end of the paddle axis (3) then transmits this position 
to the scale (6) and the optional electronic evaluators 
through the magnet tracking indicator (5) system. This 
happens safely and without packing glands.    
The flow rates shown on the scale only apply to the 
calibrated medium or to a medium with the same physical 
characteristics. 

3.2 

System design 

The TSK flowmeter mainly consists of the fitting with the 
measuri

ng chamber, the measuring elements “torsion spring” 

and “paddle,” the magnet tracking indicator system and the 
indicator unit.  
The flowmeter is manufactured as a sandwich-type 
construction and available for the nominal sizes from DN 50 
to DN 400. 
 

 

 

Fitting 

Paddle 

Paddle axis with torsion spring 

Ring-type permanent magnet 

Magnet tracking indicator system 

Scale 

Cover with window 

Base plate of indicator housing 

Coupling chamber 

Input 

4.1 

Measured variable 

 
Volume flow 

4.2 

Measuring range (lower-range and upper-
range value) 

Measuring range span 

15

–100% 

Smallest measuring range 

1.5

–10 m3/h water 

Largest measuring range 

185

–1300 m3/h water 

Electrical signal outputs  

20

–100% 

 

Summary of Contents for Heinrichs TSK

Page 1: ...Paddle Type Flowmeter for Liquids TSK Installation and Operating Instructions ...

Page 2: ...3W2 angle of rotation transmitter 6 6 CHARACTERISTIC VALUES 6 6 1 Accuracy 6 6 1 1 Reference conditions 6 6 1 2 Measured error 6 6 1 3 Repeatability 6 6 2 Influence of ambient temperature 6 6 3 Influence of fluid temperature 6 7 CONDITIONS OF USE 6 7 1 Installation conditions 6 7 1 1 Mounting start up 7 7 1 2 Device settings 7 7 1 3 Adjusting the limit transducer 7 7 1 4 Operation in hazardous are...

Page 3: ...r KINAX 3W2 transmitter with 4 20 mA output 2 wires 17 9 6 Wiring diagram for KINAX 3W2 transmitter with 4 20 mA output 3 wires 17 9 7 Wiring diagram for KINAX 3W2 transmitter with 4 20 mA output 4 wires 18 9 8 Wiring diagram for KEM 1 and KEM 2 double throw microswitches 18 10 INDICATOR UNIT 19 11 AUXILIARY POWER 19 12 CE MARK 19 13 ORDER INFORMATION 19 13 1 Available accessories 19 14 STANDARDS ...

Page 4: ...rrent flow rate in volume or mass per unit in time Applications flow measurement dosing monitoring adjusting and control of liquid and gaseous products The meter s design makes it ideal for processes under difficult and adverse operating conditions The devices are available with additional electrical equipment for process monitoring and control 3 Operational mode and system design 3 1 Measuring pr...

Page 5: ...oswitches whose switching point is activated by a cam plate KEM 1 1 Double throw microswitch KEM 2 2 Double throw microswitches Maximum make break capacity 230 VAC 50 60Hz 6 A 24 VDC 0 5 A 110 VDC 0 2 A 5 2 Analog output with the ES magneto electric transmitter The magneto electric transmitter is factory calibrated to the scale values upon shipment The signal output is supplied exclusively in a tw...

Page 6: ...ges in viscosity cause a non linear display fault 7 Conditions of use The VDI VDE guidelines 3513 Sheet 3 must be observed The meter is suitable for Liquids with sufficient flowability that are free of solids do not bond and do not tend to settle For paddle type devices viscosity factors must not be considered physically They do not play any role or play an insignificant role when the flow rates a...

Page 7: ...ions The gaskets must be suitable for the operating pressure the temperature and the measured medium With PTFE and PP coated devices use gaskets whose interior and exterior diameter correspond to the sealing strip of the device Tighten the flange bolts crosswise so that the process connections are tight Pay particular attention to the tightening torques of screws especially with PTFE and PP coated...

Page 8: ... T6 T4 7 1 4 2 2 Marking for the device when the ES magneto electric transmitter is built in DMT 00 ATEX 075 II2G Eex ia IIC T6 7 1 4 2 3 Marking for the device when the KINAX 3W2 angle of rotation transmitter is built in PTB 97 ATEX 2271 II 2G Eex ia IIC T6 7 1 4 3 Atmospheric conditions In accordance with EN 1127 a potentially explosive atmosphere is defined as a mixture of air and combustible g...

Page 9: ... 40 C or rises above 200 C The effect of the fluid temperature on the indicator unit and its built in components must be considered 7 3 2 Diagram Max ambient temperature based on the fluid temperature for the built in sensor Type ES 20 25 30 35 40 45 50 55 60 65 70 70 80 90 100 110 120 130 140 150 160 170 180 190 200 fluid temperature C Max ambient temperature C 7 3 3 Fluid pressure limit TSK1 C D...

Page 10: ...ing stainless steel Paddle axis spring stainless steel Indicator unit aluminum and safety glass window TSK1 K Fitting PP Paddle axis spring stainless steel option Hastelloy Indicator unit aluminum and safety glass window TSK1 P Fitting PTFE Paddle axis spring Hastelloy option titanium Indicator unit aluminum and safety glass window 8 4 Dimension table for TSK1 C and TSK1 S The corresponding dimens...

Page 11: ...chs Messtechnik TSK1 Installation and Operating Instructions Page 11 of 27 8 5 Dimension drawings 8 5 1 Aluminum indicator unit Flow measurement from left to the right Flow measurement from right to the left ...

Page 12: ...Heinrichs Messtechnik TSK1 Installation and Operating Instructions Page 12 of 27 Direction of flow from the top to the bottom Direction of flow from the bottom to the top ...

Page 13: ...Heinrichs Messtechnik TSK1 Installation and Operating Instructions Page 13 of 27 8 5 2 Stainless steel indicator unit Flow measurement from left to the right Flow measurement from right to the left ...

Page 14: ...Heinrichs Messtechnik TSK1 Installation and Operating Instructions Page 14 of 27 Direction of flow from the top to the bottom Direction of flow from the bottom to the top ...

Page 15: ...r remove the indicator cover insert the connector cable into the cable gland and attach it to the terminals according to the terminal diagram Tighten the cable gland securely remount the indicator cover and close it tightly 9 1 Wiring diagram for ES transmitter 9 2 Wiring diagram for ES transmitter with 4 20 mA output and 2 limit transducers ...

Page 16: ...esstechnik TSK1 Installation and Operating Instructions Page 16 of 27 9 3 Wiring diagram for ES transmitter with 4 20 mA output pulse output and limit transducer 9 4 Wiring diagram for inductive limit transducers ...

Page 17: ...sstechnik TSK1 Installation and Operating Instructions Page 17 of 27 9 5 Wiring diagram for KINAX 3W2 transmitter with 4 20 mA output 2 wires 9 6 Wiring diagram for KINAX 3W2 transmitter with 4 20 mA output 3 wires ...

Page 18: ... Messtechnik TSK1 Installation and Operating Instructions Page 18 of 27 9 7 Wiring diagram for KINAX 3W2 transmitter with 4 20 mA output 4 wires 9 8 Wiring diagram for KEM 1 and KEM 2 double throw microswitches ...

Page 19: ...the device in hazardous areas follow the applicable national installation rules 15 2 Installation start up and operating personnel Only trained specialists authorized by the system operator may carry out the installation electrical installations startup maintenance and operation They must read and understand the operating manual and follow its instructions The required mounting electrical installa...

Page 20: ...ot move If the pointer is unable to move further send the device to the head office for servicing Electrical equipment are not functioning Check the auxiliary power Are suitable power supply equipment connected have the terminals been selected correctly has the parameterization carried out correctly 19 Returning devices for repair and service Note In accordance with the applicable German waste dis...

Page 21: ...g the measured medium Since this measured medium is dangerous in water poisonous corrosive flammable we have checked that all hollow spaces of the device are free of these materials neutralized and flushed all hollow spaces of the device cross out what is not applicable We hereby confirm that in resending the device no danger to persons or the environment is posed by the residual measured substanc...

Page 22: ...Heinrichs Messtechnik TSK1 Installation and Operating Instructions Page 22 of 27 22 EC Type Examination Certificates ...

Page 23: ...Heinrichs Messtechnik TSK1 Installation and Operating Instructions Page 23 of 27 ...

Page 24: ...Heinrichs Messtechnik TSK1 Installation and Operating Instructions Page 24 of 27 ...

Page 25: ...Heinrichs Messtechnik TSK1 Installation and Operating Instructions Page 25 of 27 ...

Page 26: ...Heinrichs Messtechnik TSK1 Installation and Operating Instructions Page 26 of 27 ...

Page 27: ...Heinrichs Messtechnik TSK Installation and Operating Instructions Seite 27 von 27 23 Sales representatives Internet http www heinrichs eu 24 Notes ...

Reviews: