18 Trouble shooting
-
Indicator window clouds over:
Water in the indicator
unit.
-
Indicator cover is not tight enough:
Adjust the cover
seal, tighten the cover.
-
Window is opaque:
Corrosive atmosphere, ventilate.
-
Window ices over due to cold and damp
atmosphere:
The device can be equipped at the
factory with an air/nitrogen flush.
-
Window ices over due to very cold medium and
damp atmosphere:
The device can be equipped at
the factory with a pulled-forward indicator unit.
-
Device shows incorrect values
:
Compare process
data, density, viscosity, temperature and pressure with
the values on the scale. If they deviate, convert the
scale values using the VDE/VDI 3513 standard; for the
signal output 2 option, carry out reparameterization.
-
Pointer does not react in spite of varying flow:
The
pointer may have gotten stuck; remove the cover and
move the pointer; if the pointer can be moved easily,
the measuring body cannot move. If the pointer is
unable to move further, send the device to the head
office for servicing.
-
Electrical equipment are not functioning:
Check the
auxiliary power. Are suitable power supply equipment
connected, have the terminals been selected correctly,
has the parameterization carried out correctly?
19 Returning devices for repair and service
Note:
In accordance with the applicable German waste
disposal legislation, the owner/client is responsible for the
disposal of special waste and hazardous materials.
Consequently, all devices sent to us for repair must be free
of any hazardous materials. This also applies to possible
hollow spaces and fissures in the devices. If repair is
necessary, confirm the above-mentioned item in writing
(please use the form in the Appendix)
.
If hazardous materials remain in or on the device after it has
been returned, Heinrichs Messtechnik shall be authorized to
remove them at the client’s expense without further inquiry.
20 Replacement parts
The following parts can be ordered as replacement parts:
1 ) Indicator cover with window/gasket/screws
2 ) Scale with standard scaling
3 ) Pointer
4 ) Limit value indicator
5 ) Pointer stop
6 ) Limit value initiator
Summary of Contents for Heinrichs TSK
Page 1: ...Paddle Type Flowmeter for Liquids TSK Installation and Operating Instructions ...
Page 23: ...Heinrichs Messtechnik TSK1 Installation and Operating Instructions Page 23 of 27 ...
Page 24: ...Heinrichs Messtechnik TSK1 Installation and Operating Instructions Page 24 of 27 ...
Page 25: ...Heinrichs Messtechnik TSK1 Installation and Operating Instructions Page 25 of 27 ...
Page 26: ...Heinrichs Messtechnik TSK1 Installation and Operating Instructions Page 26 of 27 ...