ZED-D
page 18
ZED-D K11/0522
RELAY S1
Menu Item
Parameter / Function
Explanation / Values / Other
s1Funct
Relay1 Function selection
Flow: Monitoring of an adjustable flow value (s1SPunkt).
GM Limit: Monitoring of an adjustable total volume (s1TA-Lim).
PumpCtrl: Dosage pump control.
FineDose: Fine dosage valve control.
see
8.0 Relay functions
s1Char.
Relay1 Switch characteristic
Limit: Monitoring a reading (s1Spunkt).
Window: Monitoring an adjustable measuring range
(s1Spunkt...s1WPoint).
(only if
s1Funct.
is set to
Flow
)
s1SPunkt
Relay1 Switchpoint
Reading is in the same units as the flow indicator.
(only if
s1Funct.
is set to
Flow
)
s1Hyste
Relay1 Hysteresis
Reading is in the same units as the flow indicator.
(only if
s1Funct.
is set to
Flow
)
s1WPoint
Relay1 Windowpoint
Reading is in the same units as the flow indicator.
(only if
s1Funct.
is set to
Flow
und
s1Char.
auf
Window
)
s1SDelay.
Relay1 Switch delay
Value input is always in the following format
Range:_ 0,0...99,9 sec
(only if
s1Funct.
is set to
Flow
)
s1RDelay.
Relay1 Reset delay
Value input is always in the following format
Range:_ 0.0...99.9 sec
(only if
s1Funct.
is set to
Flow
)
s1TA-Lim
Relay1 Total volume limit value
Reading is in the same units as the total volume unit in
dosage unit menu group.
(only if
s1Funct.
Is set to
GM Limit
)
s1PumpAT
Relay1 Pump control lead time
Reading determines pump lead time (Rel 1) before dosage begins
Range = _0.0...99.9 sec
(only if
s1Funct.
is set to
PumpCtrl
)
see
8.2 Pump control
s1PumpDT
Relay1 Pump control coast down
time
Reading determines pump coast down time (Rel 1) after end of dosage
Range = _0,0...99.9 sec
(only if
s1Funct.
is set to
PumpCtrl
)
see
8.2 Pump control
Summary of Contents for DF-DL Series
Page 1: ...Operating Instructions for Dosing Electronic Model ZED D or DF DLxxx or Model GxxR...
Page 13: ...ZED D ZED D K11 0522 page 13 7 4 General Settings...
Page 15: ...ZED D ZED D K11 0522 page 15 7 5 Flow and analogue output...
Page 17: ...ZED D ZED D K11 0522 page 17 7 6 Relay output...
Page 19: ...ZED D ZED D K11 0522 page 19 7 7 Dosing...
Page 21: ...ZED D ZED D K11 0522 page 21 7 8 User alignment and Service Settings...