FLOW
Menu
Item
Parameter / Function
Explanation / Values / Other
fUnit
Unit of flow indicator
mL/s, mL/m, L/s, L/m, L/h, m
3
/m, m
3
/h,
GPM, GPH, UU/s, UU/m, UU/h
fDamping
Attenuation of reading fluctuations in the flow indicator
The attenuation pacifies the flow indicator.
The attenuation value is the approximate equivalent of the setting time
of the display value to c. 90 % of a measured value jump in seconds.
(Parameter is blocked at DF-...DLxxx devices).
ANALOGUE OUTPUT
Menu Item
Parameter / Function
Explanation / Values / Other
aChoice
Selection of analogue source
Flow: gauged analogue value of flow measurement
Volume: current volume in percentage of dosage
aLIFE 0
Select Life Zero
Offset at power output: 0 mA or 4 mA
at 0-10 V
≙
0 mA
→
0 V and 4 mA
→ 2 V
aLowFlow
Flow reading at 0/4 mA or 0/2 V
Lower flow reading of gauged output range,
value has the same unit as the flow indicator,
(only if
aChoice
is set to
Flow
)
aHighFlo
Flow reading at 20 mA or 10 V
Upper flow reading of gauged output range,
value has the same unit as the flow indicator,
(only if
aChoice
is set to
Flow
)
aLoVaDos
Percentage the dosage at 0/4 mA or 0/2
V
Lower volume value of gauged output range.
- Range: 0...120 %
- The difference |aHiVaDos – aLoVaDos| may not be less than 10 %.
- aLoVaDos > aHiVaDos is also possible.
(only if
aChoice
is set to
Volume
)
aHiVaDos
Percentage the dosage at 20 mA or 10 V
Upper volume reading of gauged output range.
- Range: 0...120 %
- The difference |aHiVaDos – aLoVaDos| may not be less than 10 %.
- aLoVaDos > aHiVaDos is also possible.
(only if
aChoice
is set to
Volume
)
Summary of Contents for DF-DL Series
Page 1: ...Operating Instructions for Dosing Electronic Model ZED D or DF DLxxx or Model GxxR...
Page 13: ...ZED D ZED D K11 0522 page 13 7 4 General Settings...
Page 15: ...ZED D ZED D K11 0522 page 15 7 5 Flow and analogue output...
Page 17: ...ZED D ZED D K11 0522 page 17 7 6 Relay output...
Page 19: ...ZED D ZED D K11 0522 page 19 7 7 Dosing...
Page 21: ...ZED D ZED D K11 0522 page 21 7 8 User alignment and Service Settings...