www.grupotemper.com 5
KTIR-1050
Non-Contact Infrared Thermometers
z
Do not allow the laser light beam impinge on any gas
which can
explode.
Distance & Spot Size
As the distance (D) from the object increases, the spot size
(S) of the area measured by the unit becomes larger. The
relationship between distance and spot size for each unit is
listed below. The focal point for each unit is 914mm (36”).
The spot sizes indicate 90% encircled energy.
Fig: 1
MODEL:8855 D:S=30:1
Sensor Beam
D
mm
Distance(D) to Spot Size(S)
inch
16mm
Laser Beam
S
0.4@12
1.2@36
2.0@60
10@300 30@900
50@1500