
27
Instructions manual |
www.grupotemper.com
KMD 05
Digital multimeter
readings.
To clean input jacks, follow the following steps:
1. Turn off the instrument and remove test leads.
2. Clear any dirt or other particles on tje input jacks.
3. Use a cotton ball/swab with a lubricant (i.e. WD-40) to clean off the
contacts of the input jacks.
4. Use a separate cotton ball/swab for each jack to prevent cross-
contamination.
Replacing the battery
Warning
To avoid false readings and potential dangerous situations, replace te
battery immediately when the
symbol appears.
Turn off the meter and disconnect the test leads before opening the
battery cover to prevent electrical shock or personal injury.
Use the following steps to replace the battery:
1. Turn off the meter.
2. Remove test leads.
3. Unscrew and remove battery cover from back of the meter.
4. Replace used battery with a new 9V battery.
5. Replace battery cover and fasten securely.
Replacing the fuse
Warning
Turn off the meter and disconnect test leads before opening back cover to
avoid electrical shock and personal injury.
Use the following steps to replace the fuse:
1. Turn off the meter an remove test leads.
2. Remove outer holster.
3. Unscrew and remove back cover from the meter.
4. Replace blown fuses with same amp/voltage ratings.
5. Replace back cover and fasten securely.
6. Replace outer holster.
Summary of Contents for KMD 05
Page 1: ...KMD 05 Multímetro digital Digital multimeter www grupotemper com ...
Page 28: ...28 Instructions manual www grupotemper com KMD 05 Digital multimeter ...
Page 29: ...29 Instructions manual www grupotemper com KMD 05 Digital multimeter ...
Page 30: ...30 Instructions manual www grupotemper com KMD 05 Digital multimeter ...