• Install KMD 01 communication software and USB driver in PC (see PC
software operation manual in the attached CD for details).
• Use USB cable to connect the meter and PC. The meter will display
,
and begin to send data.
• Open installed KMD 01 communication software in the PC. Data measured
by the meter can be uploaded to PC for further analysis.
NCV (non-contact Voltage Detection)
• Turn the meter to NCV grade
• Place the meter top close to the conductor.
When test voltage is greater than 110 Vac (RMS), the meter induction
voltage indicator willturn on and buzzer will give dripping alarm sound.
Note:
1. Even there is no indication, voltage may still exist.
Don’t use non-contact voltage detector to judge whether there is
voltage in the wire. Detection operation could be affected by socket
design, insulation thickness, type and other factors.
2. When inputting voltage on the meter input terminal, due to the existence
of the induced voltage, voltage induction indicator also may light.
3. Interference sources in the external environment (such as flashlight,
motor, etc.) may trigger erroneous non-contact voltage detection.
AC Voltage/DC Voltage Measurement
Voltage is the potential difference between two points. AC voltage polarity
changes over time, while DC voltage polarity does not change over time.
DC voltage measuring range of this meter: 660.0mV, 6.600V, 66.00V, 660.0V;
AC voltage measuring range of this meter: 660mV, 6.600V, 66.00V, 660.0V. The
660mV measuring range can be entered only in the manual measuring range
mode.
To measure AC and DC voltage:
• Rotate switch to voltage position.
• Respectively connect black and red test probes to COM input socket and V
input socket.
• Measure the voltage of circuit to be tested with other two ends of test
probes. (Connected with the circuit to be tested in parallel)
• Read the measuring voltage value from LCD display. When measuring DC