Step 3: adjusting the full scale value:
Your action
Message and reaction
from A 20230
Hit the front button twice.
The yellow LED double-fl ashes re-
peatedly. The currently connected
resistance is set as end value.
Now press the front button within
30 sec and hold it until the LED
lights up and then goes out
(approx. 3 sec).
Yellow LED lights up, full scale
value (end value) is saved.
Step 4:
The data are stored in the teach-in confi guration and can be activated
with the following switch positions:
Saving adjusted values,
teach-in confi guration cannot be modifi ed afterwards:
All DIP switches = 0
All rotary switches = 0
Saving the teach-in confi guration but keeping it modifi able by a
new teach-in procedure:
All DIP switches = 1
All rotary switches = 0
NOTICE!
If you do not activate the teach-in confi guration after having termi-
nated the teach-in process, the transmitter uses the confi guration
adjusted by the DIP/rotary switches.
Important note:
Cover the switches with the included self-adhesive polyimide tape
after completion of confi guration.
Teach-In Function
Summary of Contents for SensoTrans R A 20230P0 Series
Page 14: ...14 DIP Schalter ON Konfigurierung über Schalter Funktionsübersicht ...
Page 26: ......
Page 27: ......
Page 29: ...The Art of Measuring www knick de English 29 SensoTrans R A 20230P0 Resistance Transmitters ...
Page 42: ...42 DIP switch ON Configuration using Switches Overview of Functions ...
Page 54: ......
Page 55: ......
Page 70: ...70 switch DIP ON Configuration via les switches Aperçu des fonctions ...
Page 82: ......
Page 83: ......
Page 85: ...The Art of Measuring ...
Page 98: ...98 Chaves DIP ligadas ON Configuração Usando Chaves Sinopse de Funções ...
Page 110: ...110 ...
Page 111: ...111 ...