Diaphragm Pump NMP 03
Technical data
KNF Micro BA 309215-309218_NMP03-EN-0916_Rev01
Translation of original Operating and Installation Instructions, English
6
4. Technical data
General parameters
Parameter
Value
Allowed ambient temperature (°C)
- 20 to + 60
Allowed medium temperature (°C)
- 20 to + 60
Allowed storage temperature (°C)
- 40 to + 70
Tab. 3
Pump Materials of gas-contacting parts
Type designation NMP 03
KP ..
Components
Material
1)
Head plate, intermediate plate
PPS
Valve plate
EPDM
Diaphragm
EPDM
Tab. 4
DC stands for NMP 03 KP DC ..
Motor specification
S
M
L
Supply voltage[V]
6)
2.5
1.3
2.0
Max power consumption [mW]
180
160
120
Connection[-]
solder tail
wire
Protection class [-]
IP 40
IP 40
IP 40
Weight
2)
[g]
10.8
13.9
13.1
Delivery rate [ml/min]
3), 4)
330
300
320
Max. operating pressure [mbar
rel.]
3), 4))
370
350
350
Max. vacuum [mbar rel.]
3), 4)
-400
-350
-400
Tab.5
DCB stands for NMP 03 KPDC B1 / B3 (brushless)
B1: Brushless motor without electronic
B3: Brushless motor with external electronic
Motor Variants
B1
B3
Supply voltage[V]
6)
3.3
2-5
Max. power consumtion [mW]
-
155
5
)
Connection [-]
Flat flex
Pin
Protection class [-]
IP 00
Weight
2)
[g]
10.9
12.7
Delivery rate 3.3V [ml/min]
3), 4)
300
Delivery rate 5.0V [ml/min]
3), 4)
500
Max. operating pressure [mbar rel.]
3), 4)
370
Max. vacuum [mbar rel.]
3), 4)
-400
Tab.6
1)
According DIN ISO 1629 and 1043.1
2)
The weight may differ slightly from the stated value, depending on the version.
3
)
Depending on the application level (MASL), ambient temperature, material design
of the pump head and the connection hoses used, the flow rates may vary from the
values shown.
4)
Measured with air at 20°C / at atmospheric pressure
5)
At 3.3VDC control voltage
6)
Nominal use