Mini Laboratory pumps N 86.18, N 811.18
Servicing
Translation of original Operating Instructions, English, KNF 121435-121438 04/18
17
8.
Servicing
8.1. Servicing Schedule
Component
Servicing interval
Pump
Regular inspection for external damage
or leaks
Filter (Accessory)
Replace if it is dirty
Diaphragm and valve
plates/sealings (valve
plates)
Replace at the latest, when pump
output decreases
Tab. 9
8.2. Cleaning
When cleaning, make sure that no liquids enter the inside of
the housing.
8.2.1. Flushing Pump
Before switching off the pump, flush it with air (if neccesary for
safety reasons: with an inert gas) for about five minutes under
atmospheric conditions (ambient pressure).
8.2.2. Cleaning Pump
As far as possible, clean the parts with a dry cloth.
Only use solvents for cleaning if the head materials cannot be
attacked (check the resistance of the material!).
If compressed air is available, blow out the components.
8.3. Changing Diaphragm and Valves
8.3.1. N 86 K_.18
Pump is switched off and mains plug is removed from the
socket
Pump is clean and free of hazardous materials
Tubes removed from pump’s pneumatic inlet and outlet
Spare part/tool
Service Set (according to chapter 10)
Philips-head screwdriver No. 1
Small screwdriver, blade width 0.5 mm
Pencil
Tab. 10
Always replace diaphragm and valve plates together to main-
tain the pump performance.
Conditions
Spare part/tool
Information on procedure
DE
EN
FR
ES
IT
Summary of Contents for Laboport N811 KN.18
Page 2: ......
Page 28: ...KNF weltweit Unsere lokalen KNF Partner finden Sie unter www knf com...
Page 30: ......
Page 56: ...KNF worldwide Please find our local KNF partners at www knf com...
Page 58: ......
Page 84: ...KNF dans le monde entier Vous trouverez nos partenaires KNF locaux l adresse www knf com...
Page 86: ......
Page 112: ...KNF en el mundo Puede localizar los distribuidores KNF locales en www knf com...
Page 114: ......
Page 140: ...KNF nel mondo Potete trovare il vostro partner KNF su www knf com...