9. Troubleshooting
Disconnect the pump power supply before working on the
pump.
Make sure that the pump is de-energised.
Stroke not executed
Cause
Fault remedy
Pump not connected to power
supply
Connect pump to power supply
Power supply is not switched on
Switch on power supply
Connecting leads are damaged
Check connecting leads for damage
Check connecting lead entry on solenoid for damage
Connections or hoses are
blocked
Check connections and hoses
Remove blockage
External valve is closed or filter
is clogged.
Check external valves and filters
Incorrect interchange of pres-
sure and suction line connec-
tions
Remove pressure and suction lines and re-connect correctly
Pressure on the pressure side is
too high
Reduce the pressure on the pressure side of the pump
Electrical signal is outside the
defined range
Adjust pulse signal (see
Chapter 7
)
Tab. 9
Pump is not priming
Cause
Fault remedy
Suction side of pump not con-
nected
Connect the suction side of the pump
Liquid in the priming container is
too low
Fill priming container
Hose connections are not leak-
tight
Secure transition joints between hose and connections with
clamps or other clamping elements
System valve is closed or filter
is clogged
Open system valve
Clean filter
Particles in the pump
Contact KNF
The pump parts are not re-
sistant to the medium to be
pumped
Contact KNF
Incorrect interchange of pres-
sure and suction line connec-
tions
Remove pressure and suction lines and re-connect correctly
Tab. 10