Semi-Micro Osmometer K-7400S instructions, V3700
Using the EuroOsmo 7400 Software 33
Fig. 14:
Setup area of the main window
10.6 Setup
Area for editing general settings for measurement sequences and corre-
sponding reports.
Command
Explanation
Operator
Enter the name of the user.
Communication
Port
Select the COM port to which your osmometer
is connected.
10.7 Working parameters
Command
Explanation
Monitoring view
The entered value defines the scale of the moni-
tor by its final value in minutes. The temperature
pattern is shown continuously for the defined
period.
Measurement view
The entered value defines the scale of the mea-
surement by its final value in minutes.
Freezing Level
The initiation point FREEZE can be set within the
range from – 13.0 to – 4.0 ° C. Any decimal value
will be rounded in steps of 0.5 ° C.