5. Unpacking and setup
This chapter describes all preparatory steps prior to start-up.
Note:
To ensure structural stability, the device is not intended to be sta-
cked on or below other devices.
5.1
Location requirements
The intended use be ensured only if the requirements for ambient condi-
tions of the operating environment are met. You will find the ambient
conditions in the chapter „Technical data“ (see page 40).
Position the device on a level surface.
Protect the device against direct exposure to sunlight.
Set up the device at a location not exposed to air drafts such as air
con ditioning systems.
Do not set up the device near to other machines that cause floor
vibra tions.
Keep the the devices away from high frequency sources. High frequen-
cies may compromise measurement values.
At least 5 cm, if there is another device on one side.
At least 10 cm, if there are devices set up on both sides.
At least 15 cm to the cooler fan on the rear.
The device should not be exposed to major temperature fluctuations
during measurement.
Device defect
The device overheats at exposure to sunlight and insufficient air circula-
tion. Device failures are very likely.
Set up the device in such a way that it is protected against exposure
to direct sunlight.
Leave room for air circulation: See paragraph „space requirements“.
5.2
Unpacking the device
At the factory, all KNAUER devices are carefully packed for safe transport.
After unpacking, check all device parts and accessories for signs of dam-
age during transport and if necessary assert claims for damages immedi-
ately to the transport company.
The complete listing of consumable and spare parts can be found in sec-
tion „13. Repeat orders“ on page 42.
Check packaging for damage caused during transportation. If necessary,
put forward any claim for damages to the carrier.
Utility knife
General purpose
Space requirements
Ambient temperature
Prerequisite
Tools