Semi-Micro Osmometer K-7400S instructions, V3700
Unpacking and setup 12
Bruising danger
Damage to the device by carrying or lifting it on protruding housing
parts. The device may fall and thus cause injuries.
Lift the device only centrally on the side of the housing.
1.
Set up the package in such a way that you can read the label. Using
the utility knife, cut the adhesive tape and open the packaging.
2.
Remove the foam insert. Take out the accessory kit and the
instructions.
3.
Open the accessory kit and check the scope of delivery. In case any
parts are missing, contact the Technical Support.
4.
Clasp the device from below, lift it out of the packaging and place it
on its feet.
5.
Check the device for signs of damage that occurred during transport.
In case you notice any damage, contact the Technical Support.
6.
Place the device in its site of operation and remove protective foil.
Store packaging and keep the included packing list for repeat orders.
5.3
Power supply
Use only the enclosed power cable to connect the device to the mains
to make sure that the specifications stated in Technical Data are met. But
check beforehand to use power cables admitted for use in your count-
ry. Replace defective power cables only with accessories from KNAUER.
Detachable power cables must not be replaced with different cable types.
The maximum power input is 70 watts.
Electronic defect
Electronic hazard when using an identically constructed power adapter
from another manufacturer.
Only use spare parts and accessories from KNAUER or a company
authorized by KNAUER.
Note:
The nominal capacity of the connected devices must be maximum
50 % of the power supply to account for larger inrush currents when swit-
ching on the modules.
The electrical power supply at the installation site must be connected
directly to the nearest main power line.
The power must be free from ripple, residual current, voltage peaks
and electromagnetic interference.
The connectors for the mains voltage are grounded accordingly.
The device receives sufficient power with reserve capacity.
The device is intended for use with AC power networks of 100–240 V.
Make sure that the power plug on the rear of the device is always
accessible, so that the device can be disconnected from the power
supply.
Process
Next steps
Prerequisites
Power plug