Maintenance and care
BlueShadow Detector 50D Instructions V7659
Note:
If the cleaning of the lens does not have the desired effect, the
lens has to be replaced .
Mount the flow cell onto the detector .
8.3.3 Cleaning the light guide of a preparative
flow cell
The preparative flow cells have a rod shaped light guide instead of the
concave lens of the analytical cells .
Device has been switched off .
Power plug has been pulled .
Flow cell has been removed .
Tweezers
Allen screwdriver, size 3
1
Threaded ring
2
Cover
3
Spacer
4
Pressure bushing
5
Light guide with
seal ring
Fig. 17:
Preparative flow cell
Next steps
Prerequisite
Prerequisite
Tools
1 2
4
3
5