14
Installation
BlueShadow Pump 80P Instructions V7685
Process
1.
Set-up the package in such a way that you can read the label.
2.
Using the utility knife, cut the adhesive tape and open the
packaging.
3.
Lift the foam padding. Take out the accessories kit and the manual.
4.
Open the accessories kit and check the scope of delivery. In case
any parts are missing, contact the Technical Support.
5.
Clasp the device from below, lift it out of the packaging and place
it on its feet. Do not hold onto the front cover.
6.
Check the device for signs of damage that occurred during trans-
port. In case any parts are missing, contact the Technical Support.
7.
Set up the device at the operation site. Lift the foam padding.
Store packaging and keep the included packing list for repeat orders.
4.2 Connecting the piston backflushing
The pump is equipped with automatic piston backflushing. It increases
the service life of the seals and pistons, and removes contaminants from
the area behind the seals.
The piston backflushing function automatically flushes the rear piston
area of the pump head upon switch-on and in continuous operation.
Upon switch-on: The rear piston area of the pump head is flushed
automatically for 15 seconds.
In continuous mode: The rear piston area of the pump head is flushed
automatically every 30 minutes, for 15 seconds.
Next steps
Functional principle