Troubleshooting
BlueShadow Pump 80P Instructions V7685
System message
Solution
Cannot operate
with an empty link
The
link
is empty. First create a
link
.
Cannot read data
from FRAM
Switch the device off and on. Inform the techni-
cal support in case the system message repeats
itself.
Cannot read RTC
Switch the device off and on. Inform the techni-
cal support in case the system message repeats
itself.
Cannot start time
table
Edit the data on the device display or in the
chromatography software.
Cannot use non-ex-
isting component!
Change the
setup
settings or change the gradi-
ent in the program or in
setup
.
Cannot write data
on FRAM
Switch the device off and on. Inform the Techni-
cal Support in case the system message repeats
itself.
CRC failed
Switch the device off and on. Inform the techni-
cal support in case the system message repeats
itself.
E
Error input
activated
Device error, change device settings.
G
GUI communication
failed (internal)!
Switch the device off and on. Inform the techni-
cal support in case the system message repeats
itself.
I
Instrument in error
state
Switch the device off and on. Inform the techni-
cal support in case the system message repeats
itself.
Instrument in
standalone mode
Change the entry in the Setup menu.
Change the entry in the chromatography
software.
Instrument in
standby mode
Press the standby button to end the standby
mode.
Instrument remote
controlled
Change the entry in the Setup menu.
Change the entry in the chromatography
software.
Insufficient access
Change the entry.
Invalid command
Change the entry.
Check the cable connections.
Invalid parameter(s) Check the validity of the parameter(s).
Invalid time in time
table
Correct the time entry.