P 2.1L user manual V6840, version
3.
5
Technical Data
37
Technical Data
Solvent delivery
Pump type
Preparative HPLC pump
Delivery system
Dual-piston pump
Flow rate range
100 ml pump head:
0.1 - 100 ml/min
250 ml pump head:
0.1 - 250 ml/min
500 ml pump head:
0.1 - 500 ml/min
1000 ml pump head:
0.1 - 1000 ml/min
Maximum pressure
100 ml pump head:
400 bar to 100 ml/min
250 ml pump head:
225 bar to 100 ml/min
Linear reduction: 225-200 bar of 100-
150 ml/min
200 bar of 150-250 ml/min
500 ml pump head:
100 bar to 500 ml/min
1000 ml pump head:
75 bar to 350 ml/min
Linear reduction: 75-50 bar of 350-600 ml/
min
50 bar of 600-1000 ml/min
Flow rate accuracy
±2 % at 2 – 50 % of the flow range with metha-
nol-water mixture (10/90, v/v)
Flow rate precision
< 0,1 % RSD
Gradient support
Isocratic HPLC pump
binary or ternary LPG system
binary HPG system with additional pump
System protection
P
min
and P
max
adjustable
Communication
Control
LAN
Pin header connectors (Analog IN, Start IN,
Error IN)
Control Unit
Programming
19 programs, 9 links, WAKE UP program
Supply frequency
50 – 60 Hz
Power consumption
Pump: maximum 320 W
Binary or ternary valve block: 5 W
Protection type
IP 20
Technical parameters
Temperature range
4 – 40 °C; 39.2 – 104 °F
Air humidity
below 90 %, non-condensing