background image

Tubing connections

Legend

 

 

from eluent bottle via pump to pressure sensor

   from pressure sensor via manual injection valve (6 Port 2 Position valve) 

         to sample loop selection valves (16 Port Multiposition valves) 

 

 

from manual injection valve via column bypass valve (6 Port 2 Position 

         valve) and column to detector

  

from detector via fractionation valve (16 Port Multiposition valve) 

         to fraction bottle

   to waste bottle

After installation of single devices according to their respective manual the GPC system ist set up. Lastly, 
tubings are connected with the system as highlighted in diff erent colours.

Damage to Tefzel® capillary may occur when certain parts are used which are not suitable with Tefzel®. Do 
not use stainless steel ferrule 1/16” in combination with stainless steel fi tting 12 mm UNF10/32 on Tefzel® 
tubings, though both parts are included in the Accessory Kit for 6-Port/3-Channel Injection Valve (order 
number F1357).

Component defect

This document is subject to technical changes. Please go to www.knauer.net 
for latest version, ordering details and overview of article numbers.

Note: 

The inlet tubing of the AZURA® Pump P 4.1S included in FPGA is too short for the AZURA® GPC 

Cleanup System. Use a longer inlet tubing included in A50041. 

Note: 

In case of a fl uctuating detector signal which results from degassing in the fl ow cell, use a tubing 

with a low inner diameter of 0.25 mm between detector and fractionation valve.

4

© KNAUER

All rights reserved.

Version 2.0

Last update: 2017-09-27

KNAUER Wissenschaftliche Geräte GmbH

Hegauer Weg 38

14163 Berlin, Germany

Tel.: +49-(0)30-809727-0

Fax: +49-(0)30-8015010

E-mail: [email protected]

Internet: www.knauer.net

Note: 

Depending on your preferences, the manual valve and electrical 6 Port 2 Position valve can be 

exchanged to function as column bypassing/selection and sample injection.

Reviews: