Kleinwaechter EFM 023 BGT Manual Download Page 9

 

Maintenance 

Display 

The display is a double-spaced alphanumeric LCD display with 12 digits in each row (2 x 12). In 
the upper line the selected distance is displayed in cm. In the bottom line the measured charge is 
displayed in Volt.  Over 999V on the display changes to kV display.  

The displayed value is always triple-digit: 

 
Examples: 

 

 

If the measure field strength is too high „overflow! “ is displayed, the distance has to be 
increased. 

 

Battery Monitoring 

The device has permanent battery voltage monitoring. By undercutting a battery voltage of 7.6V 
the upper line displays: „Low Battery!“. In that case the 9V battery has to be charged or renewed. 
By undercutting 7.1V the display shows “auto off” for about 2s and the device powers off to pre-
vent further damage.   

Attention:

 

 

Use only Rechargeable 9V – NiMH Battery or an 9V Alkaline Battery 

Batteries have to be charged outside the device in a suited battery charger. Please pay attention 
to the manufacturer instructions! Used batteries can be returned to us or have to be disposed 
professionally.  

 

Grounding 

The device has to be grounded properly to measure the amount and the polarity of an electro-
static field. For that purpose the device has to be connected to ground by using the grounding 
jack (4). Grounding the device by a grounded person (e.g. over conductive shoes or wrist strap) 
touching the grounding jack suffices under most circumstances. 

Maintenance 

The modulator system or parts of it must not be touched. Those parts have to be protected of 
dust, isolating impurity layers, color- or paint haze and condensate.  

If required the modulator system can be cleaned with ethyl alcohol and a lint-free cotton cloth.  

 

 

578V 

3,85kV 

24,0kV 

274kV 

Summary of Contents for EFM 023 BGT

Page 1: ...MANUAL EFM 023 BGT System Prof Dr Ing Hans Kleinw chter Walking Test with EFM 023 and UAC 110...

Page 2: ...nt ranges 7 E Field Meter Measurement 7 Max Value 7 Analog Output 7 Changing the measurement distance 8 Spacers 8 Display 9 Battery Monitoring 9 Grounding 9 Maintenance 9 Battery Replacement 10 Zero P...

Page 3: ...aterial that releases electrons is charged positive The material admitting electrons is charged negative There are different ways to avoid or to discharge electrostatic charges But to find an effectiv...

Page 4: ...esolution 10Bit Battery 9V alkaline or NiMH battery Operating time Approx 10h with alkaline battery Measuring range Voltmeter Distance Measuring range Max resolution 1cm 0 10kV 1V 2cm 0 20kV 2V 5cm 0...

Page 5: ...M023 In the middle of the left side here is the output socket for the analog output 1 Cover cap 5 Spacer 2 Modulator system 6 Zero point trimmer 3 Button function on 7 Alphanumeric LCD display 4 Groun...

Page 6: ...briefly to change to a new distance or to CPS mode wait approx 2s to select the displayed function Aus Press function on 2 times in quick succession Initiation By pressing the function on button on t...

Page 7: ...Field Meter Mode In this Mode it s possible to select the following ranges 20KV m 200kV m or 1MV m To change the ranges press the function on key briefly The range changed as following In this Mode t...

Page 8: ...m the plate behind the modulator blade The plate is located 6 4mm in front of the devices front plane After switching the device off and on always 2cm distance are selected Without the manual E Field...

Page 9: ...further damage Attention Use only Rechargeable 9V NiMH Battery or an 9V Alkaline Battery Batteries have to be charged outside the device in a suited battery charger Please pay attention to the manufac...

Page 10: ...of the device The guarantee expires if the device was opened Batteries and accumulators are not allowed in the household waste Every consumer must by law dispose of all batteries and accumulators at...

Page 11: ...oltage measuring head MK 023 you get a high impedance volt meter with an impedance off 1016 Ohm That takes to measure static voltage at isolated conductive parts up to 4kV Select distance 1cm that wil...

Page 12: ...Alphanumeric LCD display 4 Metal spacers 5 Measuring head MK023 Specifications Voltage measuring head Dimensions L x W x H Approx 80mm x 40mm x 37mm Weight Approx 80g Measuring voltage Approx 4kV Han...

Page 13: ...Switch on EFM 023 and wait until Dist 2cm displayed Connect CPS socket to ground connect Hand electrode with Silicon cable to the measure input screw theVoltage measuring head 5 on the spacers replace...

Page 14: ...Start the walking test by pressing the Start button in the software End the walking test by pressing the Stop button By pressing the Reset button max Values will be erased Scope of Delivery The basic...

Page 15: ...s on the modulator system must be avoided The use of the device in power plants or comparable areas is prohibited The device can not measure alternating fields 1Hz Ad On facilities Audit Kit 1 CPM Ord...

Page 16: ...A Ohm Meter TOM 100 with Temperature and humidity Sensor TF 100 Additional Scope of Delivery Plate Electrode 75mm x 150mm 20pF HV Converter 1 100V LC TERA Ohm Meter TOM 100 TOM Surface Measure Electro...

Reviews: