![Kleinwaechter EFM 023 BGT Manual Download Page 14](http://html1.mh-extra.com/html/kleinwaechter/efm-023-bgt/efm-023-bgt_manual_1977518014.webp)
14
Manual EFM 023 BGT
Walking test with UAC 110 on a PC :
Prepare EFM 023 as description before, but select => «E-Field-Mode» instead of «MAX-Mode»
The displayed ranges correlatively now :
±20kV/m
±200V
±200kV/m
± 2kV
±1MV/m
± 10kV
Attention !!
The max. measuring voltage value is 4kV
.
Connect UAC 110 into a free USB – interface on your PC and wait until the driver is installed
Start the „KL_ReadOut” Software.
By Device => EFM 023 => Voltmeter(MK22) => Range select the same range as selected at the
EFM 023.
Select with View => Chart or => Display
Start the walking test by pressing the Start button in the software. End the walking test by
pressing the Stop button.
By pressing the Reset button max. Values will be erased.
Scope of Delivery
The basic equipment of the electrostatic field meter includes the following components:
•
ESD transport case with conductive foam interior
•
Electrostatic field meter
•
2× 9V NiMH accumulators
•
Plug-in charger
•
Grounding spiral cable with alligator clip
•
conductive CPS socket
•
Measuring head MK 023
•
Hand electrode d20mm x 100mm
•
Silicon cable 3m
•
Analog Converter UAC 110
•
Connecting cable BNC
•
USB Y-cable
•
KL_ReadOut Software
•
Manual
•
Certificate of calibration
Calibration
A check of the measurement values is recommended annually.