OPERATING INSTRUCTIONS
DIODE TEST
1. Insert the RED test lead into the VΩ jack
5
and the BLACK
test lead into the COM jack
4
, and rotate the Function selector
switch
2
to the Continuity/Resistance/Diode test
setting.
NOTE:
The meter will default to continuity testing in this mode.
Press the 'SEL' button
9
twice to enter Diode testing mode.
The Diode icon will appear on the display.
2. Touch the test leads to the diode. A reading of 200-800 mV on
the display indicates forward bias. '
OL
' indicates reverse bias.
An open device will show '
OL
' in both polarities. A shorted
device will show approximately 0 mV.
FREQUENCY/DUTY CYCLE
1. Insert the RED test lead into the VΩ jack
5
and the BLACK
test lead into the COM jack
4
, and rotate the Function selector
switch
2
to the Frequency/Duty cycle
setting.
NOTE:
The meter will default to frequency testing in this mode. To
enter Duty-cycle testing mode, press the 'SEL' button
9
once. Ensure
that the appropriate icon (either
Hz
or
%
) appears on the display.
2. Measure by connecting the test leads across the circuit.
Black lead Red lead
Black lead Red lead
CAPACITANCE
1. Insert the RED test lead into the VΩ jack
5
and the BLACK
test lead into the COM jack
4
, and rotate the Function selector
switch
2
to the Capacitance setting.