Shock Accelerometer
Page 9
002-096e-01.19
Outline dimensions used in mounting that vary by series designation
are shown in Figure 3. Type designations that include the letter E are
sized in metric units.
Surface Dia.
D
L1 inch [mm]
L2 inch [mm}
3/8” minimum
10-32 UNF
0.156 [4.0]
0.315 [8.0]
1/2” minimum
¼-28 UNF
0.22 [5.6]
0.42 [10.7]
Table 3:
Mounting Surface Preparation
6.4
Securing Cables
Figures 4 and 5 show the correct and incorrect ways for installing
cables. Allow a sufficient radius to ensure a proper strain relief. The
actual radius will depend on the cables being used. It is recommended
that cables be secured to the accelerometer mounting surface to
minimize cable and connector fatigue failures. Secure cables with a
cable clamp.
Figure 4:
Correct Cable Strain Relief