Maintenance
Interval
Maintenance work
At least every 6 months
Chapter 9.1 “Safety inspection” on page 53
.
9.1 Safety inspection
Perform safety inspection every two years!
The unit should be inspected by the owner in line
with DGUV regulation 3 (formerly BGV A3).
Subject the unit to a safety inspection at regular
intervals, but at least every two years.
The safety inspection performed by the owner contains the fol-
lowing individual inspections and their documentation:
1 - Visual inspection
2 - Function check
3 - Temperature check
4 - Temperature alarm test
5 - Inspection of the condenser
NOTICE!
If there is any doubt that the unit is in the correct
condition, immediately stop using the unit. To pre-
vent unintended use, label the unit accordingly.
Contact the service department
.
1.
Check the entire unit for completeness, correct set-up and
damage.
2.
Check the following parts of the unit individually for damage:
n
Door handle
n
Interior
n
Door seal
3.
If there is damage and functionality is not ensured, decom-
mission the unit and contact the service department
Contents of the safety inspection
Visual inspection
Maintenance
Safety inspection
27.05.2021
Refrigerator/freezer LABO series
53