![Kipp & Zonen Brewer Mk III Operator'S Manual Download Page 54](http://html1.mh-extra.com/html/kipp-and-zonen/brewer-mk-iii/brewer-mk-iii_operators-manual_1971770054.webp)
MKIII OPERATOR’S MANUAL
51
18. Current Activity
- Comments about what the Brewer is currently doing.
19. Most Recent
Measurement
429.4
- Most recent successful measurements of a certain
type (ds O3, zs O3 and ds SO2).
20. Ave. of Day’s
Measurements
430.7
- Average of all successful measurements of a certain
type taken in the current day.
A typical screen is shown below.
The lower part of the screen indicates in real time what tasks the Brewer is performing, and there is
a Status Box which shows some important status information including:
- the latest DUV measurement
- the running average O
3
values
- the latest SL “ R6 “ value
- the latest HG measurement time and temperature
- the current temperature
If the software has crashed, then an error and a program line number will be displayed on the
screen.
A glance at the screen will determine if the Brewer is operating as expected, and any discrepancies
should be noted and corrected.
END OF DAY PRINTOUT
Look at the End of Day printout (or “D” file if PD is used in ED) and check especially
the
diagnostic tests, and summary data.
5. ROUTINE OPERATIONS AND MINOR MAINTENANCE
Summary of Contents for Brewer Mk III
Page 1: ...Operators manual Spectrophotometer Brewer MkIII...
Page 3: ...ii...
Page 5: ...2...
Page 7: ...4...
Page 25: ...MKIII OPERATOR S MANUAL 22...
Page 29: ...MKIII OPERATOR S MANUAL 26 Figure 3 4 Brewer Spectrophotometer Tripod...
Page 63: ...MKIII OPERATOR S MANUAL 60...
Page 83: ...MKIII OPERATOR S MANUAL 80...
Page 91: ...MKIII OPERATOR S MANUAL 88...
Page 93: ...MKIII OPERATOR S MANUAL 90...
Page 97: ...MKIII OPERATOR S MANUAL 94...
Page 129: ...MKIII OPERATOR S MANUAL 126...
Page 135: ...MKIII OPERATOR S MANUAL 132...