MKIII OPERATOR’S MANUAL
129
Motor <number>: Motor lost.
The motor controller and the instrument controller are out of step. This error will be corrected
next time the moter is moved. If this error persists for the given motor, please notify KIPP &
ZONEN <[email protected]> of the circumstances in which this message was
generated.
Motor <number>: Move attempted outside of logical bounds
An attempt was made to move the given motor out of bounds. If this error persists, please
notify KIPP & ZONEN <[email protected]> of the circumstances in which this
message was generated.
Motor <number>: Movement attempted with too large a jerk
Please notify KIPP & ZONEN <[email protected]> of the circumstances in which
this message was generated.
Motor <number>: Movement attempted with too large an acceleration
Please notify KIPP & ZONEN <[email protected]> of the circumstances in which
this message was generated.
Motor <number>: Movement attempted with too large a velocity
If this error persists for the given motor, please notify KIPP & ZONEN
<[email protected]> of the circumstances in which this message was generated.
Motor <number>: time allotted (<ss> seconds) for motor movement has expired. Motor
stopped.
The value <ss> comes from the MOTOR.TIME.OUT configuration parameter.
New value set for watermark <nameOfWatermark>: <latestMark> <unitsOfWatermark>.
The indicated watermark value has been changed to <latestMark>. The possible watermarks
are identified in Appendix H.
No Flash config segment in Flash to copy into RAM upon initialization
Attempt to initialize when no configuration information had been loaded yet. This message is
normal as part of the process of loading new firmware.
Parallel command operation not supported:
A command was issued which requested that two operations run in parallel when that is not
legal. The subsequent two messages indicate that the command is bad and present the
command. The following commands cannot start while another command is still in operation:
I, PMT, R, SAVE, STEPS, USECONFIG
PMT counter failure
The two independent counters on the PMT did not match count rates (differed by more than
2).Counter 2 has been disabled until explicitly re-enabled or until the instrument reinitializes.
APPENDIX I FIRMWARE LOG
Summary of Contents for Brewer Mk III
Page 1: ...Operators manual Spectrophotometer Brewer MkIII...
Page 3: ...ii...
Page 5: ...2...
Page 7: ...4...
Page 25: ...MKIII OPERATOR S MANUAL 22...
Page 29: ...MKIII OPERATOR S MANUAL 26 Figure 3 4 Brewer Spectrophotometer Tripod...
Page 63: ...MKIII OPERATOR S MANUAL 60...
Page 83: ...MKIII OPERATOR S MANUAL 80...
Page 91: ...MKIII OPERATOR S MANUAL 88...
Page 93: ...MKIII OPERATOR S MANUAL 90...
Page 97: ...MKIII OPERATOR S MANUAL 94...
Page 129: ...MKIII OPERATOR S MANUAL 126...
Page 135: ...MKIII OPERATOR S MANUAL 132...