KI 9000 UM-14
25
SPECIFICATIONS
KI 9600XL Series Large Area Optical Power Meter
Detector type
Response
λ
nm
Damage
level
dBm
Calibration
λ
nm
Power
range
dBm
Midrange
linearity¹
dB
Calibration
Accuracy
2
%
Polarisation
Sensitivity
dB
Total
Uncertainty³
dB
λ
Sensitivity
5
± 30 nm
5 mm
Ge
600 ~ 1650
+15
635,650,660,780,1610,1625
850,1300,1310,1390,1490,1550
+10 ~ -35
+10 ~ -40
0.06
2 %
< 0.005
0.5
0.04
7 mm Ge
600 ~ 1650
+15
635,650,660,780,1610,1625
850,1300,1310,1390,1490,1550
+10 ~ -30
+10 ~ -40
0.06
2 %
< 0.005
0.5
0.04
5 mm Si
350 ~ 1100
+10
470, 520,
635, 650, 660,
780, 850, 980
+5 ~ -50
+5 ~ -60
0.04
2 %
< 0.005
0.3
0. 03
typical
typical
max
typical
Note 1: Mid-range linearity @ 1550 nm for InGaAs & Ge, or 850 nm for Si. Non-coherent light, with APC connector. Excludes top 5 dB and bottom 10 dB of range.
Note 2: Calibration condition: non coherent light, -35 ± 5 dBm, 23 ± 1
°
C, ± 1 nm, 10 ± 3 nm FWHM, PC ceramic connector, 100 µm fiber.
Note 3: Includes contributions of: varying optical connector types, calibration uncertainty, linearity over temperature & range, and fiber core diameter up to 1 mm ( for 5 mm detector, max NA= 0.5) or fiber core diameter up to 3 mm (for 5 mm
detector, max NA= 0.3).
Note 5: At calibration wavelengths in bold type.
Summary of Contents for KI 9600 Series
Page 1: ...OPERATION MAINTENANCE GUIDE KI 9600 Series Power Meter KI 9800 Series Light Source...
Page 5: ...KI 9000 UM 14 3 KI9600A Series Optical Power Meter and KI9800 Series Optical Light Source...
Page 50: ...KI 9000 UM 14 48 NOTES...
Page 51: ...KI 9000 UM 14 49 NOTES...
Page 52: ...KI 9000 UM 14 50 NOTES...