![Kimo DB 200 User Manual Download Page 12](http://html1.mh-extra.com/html/kimo/db-200/db-200_user-manual_1967489012.webp)
E3 :
this screen reminds the general format of the session of measurement:
•
Start :
date of measurement start
•
End :
date of measurement end (here, not entered
because measurement is still ongoing).
•
Time :
measurement time (here, not entered
because measurement is still ongoing).
•
Mode :
measurement mode type (here, L-Stock)
•
Weig. :
weighting frequency type (here, A)
•
Cst :
time weighting (here, F)
•
Lpk :
weighting frequency type of the peak sound level (here, C)
•
Samp. :
sampling rate (here, 01s)
5.2.4 Stop the measurement
We stopped the measurement manually through the pictogram.
Note :
The
manual
command (Start - Stop) has priority over the
I/O
command (even if the initial choice is a start from the
I/O
command).
The pause function is inoperative.
5.2.5 Store data
At the end of the measurement, in
manual
mode, a screen of all results is displayed. It is then possible to :
➢
Accept measurements and store them : move the cursor on
Sto.
and press once on OK key to confirm. A banner displays the
storage phase.
➢
Reject measurements and not store them : move the cursor on
RST
and press
twice
on OK key to confirm. A banner displays
the phase of suppression of data.
Make some measurements
12
Screen 3
Ongoing
measurement
Results
Storage
Results
Deletion
Ongoing
measurement
RST active